10.2.3.10
Obj par
10439203;a2
Figure 10.40
Object Parameters
dialog box
You use this command to set the object parameters
Emissivity
,
Distance
,
T Reflected
,
T Atmosphere
,
Rel humidity
,
External optics
,
Optics transmission
, and
Optics
temperature
. The parameters are selected by moving the joystick up/down and set
by moving the joystick left/right. These parameters settings will be used by all mea-
surement functions that have not been set locally.
Click
Emissivity table
to display an emissivity table on the screen. You can use this
emissivity table to find emissivities for a number of different materials. An emissivity
table can be created and edited in FLIR Systems’s PC software.
➲
Please note the following:
■
The emissivity file can be stored at root level or at directory level. However, the
camera software prioritizes files that are stored at directory level and the directory
has to be selected in order to store the emissivity file in the camera memory. If the
camera software does not find an emissivity file at directory level, it searches for
similar files at root level and saves those instead.
■
If you enter an emissivity value less than 0.30 the emissivity box will begin flashing
to remind you that this value is unusually low.
■
The transmission factor is applied to the signal and not to the temperature
For more information about object parameters, see section 18 – Thermographic
measurement techniques on page 175.
10.2.3.11
Deactivate local par.
Point to
Deactivate local par.
and press the joystick to delete all locally set parameters.
Locally set parameters are the parameters you set in
e.g.
the
Spot settings
dialog
box.
10
110
Publ. No. 1558240 Rev. a156 – ENGLISH (EN) – March 1, 2006