Operation — Operator instructions
2. The system automatically measures and registers the sample. When completed, a
message is displayed in the visitor window.
3. Repeat the two steps above until 10 healthy persons have been screened.
The
Status
indicator shows the number of registered samples.
4. When the
Status
indicator is green (OK), the system is ready for screening.
8.3.2
Screening procedure
Once the screening station is up and running, the FLIR Screen-EST Kiosk normally does
not require any action from you. Depending on your workflow, you may need to be avail-
able for guidance and alarms.
8.3.2.1
Screening procedure
This is what happens during the screening:
1. The visitors enter the queue to the screening station.
2. One visitor at the time goes to the screening position.
Note
• Removing eyeglasses is crucial for a correct screening result.
• It is important to stand at the screening position (not closer or further away from
the camera).
• The eyes must not be covered.
3. The visitor stands still and looks at the visitor display. The system automatically
measures the temperature.
#T810570; r. AA/72255/72255; en-US
27