2. eCal procedure steps:  

 

 

 

 

Descriptions 

 

 

Step 1  START eCal 

 

Command: 

@011000000001020001EB 

 

Start the eCal process. 

 

Response: 

@011000000001EE 

 

 

ACK 

 

Step 2.  Total Load cell Emax entry 

Command: 

@01100003000204000001F4F1 

Total LC is 500kg. 

Response: 

@011000030002EA 

 

 

ACK 

 
Step 3.  Average Load cell mV/V entry 

Command: 

@0110000500020400004E2F67 

LC sensitivity: 2.0015

mV/V

 

Response: 

@011000050002E8 

 

 

ACK 

 
Step 4.  Estimated Dead load entry 

Command: 

@0110000700020400000000E2 

Dead load is 0kg. 

Response: 

@011000070002E6 

 

 

ACK 

 

Step 5.  Scale Capacity entry 

Command: 

@01100009000204000000FAE6 

Scale capacity is 250kg.  

 

Response: 

@011000090002E4 

 

 

ACK 

 
Step 6.  SAVE with estimated dead load value 

Command: 

@011000000001020007E5 

 

Save the eCal values. 

 

Response: 

@011000000001EE 

 

 

ACK 

 
 

Or SAVE with automatic zero adjustment. 

 

Unload the scale and send the command below  
to start the zero adjustment. 
Command: 

@011000000001020015D7 

 

Save eCal with Zero Adj 

Response: 

@011000000001EE 

 

 

ACK 

 
Step 7.  APPLY 

Command: 

@011000000001020009E3 

 

Apply the calibration. 

Response: 

@011000000001EE 

 

 

 

ACK 

 

Step 8.  STOP eCal 

Command: 

@011000000001020002EA 

 

Stop the eCal process 

Response: 

@011000000001EE 

 

 

ACK 

 

3. Digital Filter Setting: 
 

Command: 

@011000220001020000CA   

Set the Fast filter  

Command: 

@011000220001020001C9   

Set to Medium filter  

Command: 

@011000220001020002C8   

Set to Slow filter  

Response: 

@011000220001CC   

 

ACK 

 

 

Summary of Contents for FAA-28

Page 1: ...FAA 28 Analog Amplifier Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...

Page 2: ...sioning 9 4 Adjustments 10 4 1 Adjustment with Rotary Switches 10 4 2 Fast Adjustment to Nominal Output Range 12 4 3 eCal Electronic Adjustment Via xFace Software 12 4 4 Adjustment eCal With PLC 13 4...

Page 3: ...this publication may be reproduced stored in a retrieval system or transmitted in any form or by any means mechanical photocopying recording or otherwise without the prior written permission of Flint...

Page 4: ...t the electronic calibration eCal and fast calibration with 20 max test load reduces the calibration time All analog outputs of the instruments are matched in the production to perform calibration at...

Page 5: ...to chapter 9 Features Minimized zero and span drifts due to its microcontroller technology and high accuracy very low temperature drift 24 bits ADC and 16 bits DAC converters Long time stability and l...

Page 6: ...PC Preadjusted instrument for calibration at PLC Digital Filter Programmable 9 step adjustable digital adaptive filter Analogue outputs Current output for 0 20 mA and 4 20 mA or voltage output for 0 5...

Page 7: ...nst radio frequency interference and the cabinet shall be connected to ground against the electromagnetic disturbances Load cell cable and analog output cable trays must be separated from others if po...

Page 8: ...rgizing to avoid damages of the instrument and load cells In 4 wire installations the sense and excitation pins with the same polarity should be short circuited at the connector side If you have junct...

Page 9: ...one of the analog output types can be used at the same time and has to be selected in the setup type Install the analog output measuring instrument for adjustment e g multimeter Pin Name Definition I...

Page 10: ...osition Connect or disconnect TxD and RxD pins according the table above to set the output range Turn the adjustment switch to the position After 2 seconds turn in back to the 0 position if the instru...

Page 11: ...flashes to indicate the instrument is not in operation mode Zero Adjustment Connect the measurement instrument Multimeter is recommended to the analog output Unload the scale the scale must be empty...

Page 12: ...2 5VDC Turn the adjustment switch to the G position from the analog output type position you set before Increase or decrease the analog output by adjustment rotary switch to the calculated output val...

Page 13: ...n and turn it back to 0 position 2 seconds later Turn the programming switch to analog output type position for operation For example for 4 20mA range and for 100kg capacity scale load the scale 20kg...

Page 14: ...RS232C serial interface by commands or by xFace software If you want to use your own software to calibrate the FAA 28 from your PLC or PC you may use the commands below The yellow colored bytes are th...

Page 15: ...01100009000204000000FAE6 Scale capacity is 250kg Response 011000090002E4 ACK Step 6 SAVE with estimated dead load value Command 011000000001020007E5 Save the eCal values Response 011000000001EE ACK Or...

Page 16: ...e_Add Function Last_data STX and CSUM are neglected while calculating CSUM Example For Medium 1 filter 011000220001020001XX CSUM 0 01 10 00 22 00 01 02 00 01 0 37 C9 Command is 011000220001020001C9 Ca...

Page 17: ...n the programming switch to S1 or S2 position from operation position from V or I position Turn the Adj switch to positions position and turn it back to 0 position 2 seconds later Turn the programming...

Page 18: ...ted Off 0 10VDC Open circuited Flash 0 5VDC Short circuited Blinking The status of the LEDs in the operation type is given in the table below Refer to chapter 9 in case of the Err LED turns on The ana...

Page 19: ...ntains the xFace software v2 03 or higher into the CD ROM drive Double click Setup exe to start the installation The setup Wizard will be displayed Follow the menus in the setup wizard step by step Af...

Page 20: ...oints in section Setup 9 In section Status instrument information are available 10 Power off the instrument before disconnecting from PC Bring the programming switch to the analog output type position...

Page 21: ...not let unauthorized people interfere with the instrument FRONT PANEL LEDS DEFINITION Run V I Err Off Off Off No power Board failure On On Off Operation in 4 20mA output type On Off Off Operation in...

Page 22: ...FAA 28 Technical Manual Rev 1 1 0 February 2022 Page 21 of 21 Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www flintec com...

Reviews: