14
If the temperature difference between the maximum temperature and the minimum is more than 3°C, it will
show void. That means the result of the test is a fail. You will need to run the vacuum test again after the
chamber has cooled down.
Data
The internal memory will store the information of the last 9999 cycles.
USB Flash memory (Optional)
A USB drive can be used as a method of storing a report of the cycle. To do so, insert the USB drive into the
slot located on the service door of the sterilizer.
The information will automatically output directly to the USB drive after the cycle has completed. The name
of the file is determined by the serial number of the machine and the cycle number.
For example:
The serial number is E00001. The cycle number is 0012.
The file name in the USB stick is 01001200.txt.
The first two numbers represent machine number.
The middle four numbers represent cycle number.
The last two numbers represent error code.
E.g. 00:no error; 01: error E01
Printer (Optional)
If installed, you can see the icon in the screen stop flashing.
At the end of each cycle the printer will print out a report of the cycle.
Note: If there is no paper inside the printer, the icon will flash.
Report
Internal Memory
In this menu you can get the information of all the cycles stored in the internal memory of the sterilizer.
Select “Report” from the main menu and press button, you will see the list of records.
Select the record by pressing button.
Press button to print and save the report.
Press
button to exit.
Program
Basic Set
Report
Label
00012
00011
00010
00009