7 Mounting
PIOX S72*
7.2 Transducers
39
UMPIOX_S72xV1-9EN, 2022-05-15
Step 1: Disassembly of the transducer mounting fixture Variofix L
• Disassemble the transducer mounting fixture Variofix L.
Step 2: Fixation of the clasps to the tension straps
• Select the installation instruction of the supplied clasp:
Band clamp clasp
The clasp is fixed to the tension strap (see Fig. 7.14).
Quick release clasp
The clasp is fixed to the tension strap (see Fig. 7.15).
• Cut the tension strap to length (pipe circumf at least 120 mm).
Ratchet clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
Fig. 7.13:
Disassembly of the transducer mounting fixture Variofix L
1 – cover
2 – screw
3 – nut
4 – rail
5 – tension strap clamp
Fig. 7.14:
Band clamp clasp with tension strap
Fig. 7.15:
Quick release clasp with tension
strap
Caution!
The edge of the tension strap is very sharp.
Risk of injury.
→
Debur sharp edges.
→
Wear the required personal protective equipment.
→
Observe the applicable rules.
1
2
3
4
5
5