20 Advanced functions
PIOX S72*
20.7 Event protocol
187
UMPIOX_S72xV1-9EN, 2022-05-15
20.7
Event protocol
If an error occurs, an error message will be displayed in the first line indicating the symbol
. The error message can be
displayed via the menu item
Event log
.
• Select the menu item
Event log
.
• Press ENTER.
A list is displayed containing all error messages since the last time the transmitter was switched on.
• Press
and
to select an error message.
• Press ENTER.
The display indicates the cause of the error.
Tab. 20.6: Pictograms for the status display of the event triggers
no.
Function
(switching condition)
Type
(holding behavior)
current state
R
=
1
MAX (x>limit)
Non-hold
deactivated
(false state)
2
MIN (x<limit)
Hold
activated
(true state)
3
Within range
Hold for a while
4
Out of range
ERR (x=fail)
Example
R1 =
Special functions\System settings\Event log
Notice!
After reading out the event protocol the error message symbol will be deleted on the display, even if the error has not
been eliminated yet.
The event protocol will be deleted after a restart of the transmitter.