18 Data logger
PIOX S72*
18.2 Deletion of the data logger
167
UMPIOX_S72xV1-9EN, 2022-05-15
Storage mode
• Select the menu item
Configuration
.
• Press ENTER until the menu item
Storage mode
is displayed.
• Press ENTER.
• Select
Sample
to store the current measured value.
• Select
Average
if the average of all undamped measured values of a storage interval is to be stored.
Further parameters for storing
It can be defined whether the following parameters are to be stored together with the measured values.
• Select
Yes
to store the value. Select
No
in order not to store the value.
18.2
Deletion of the data logger
• Select the menu item
Delete meas. values
.
• Press ENTER.
• Select
Yes
or
No
.
• Press ENTER.
Special functions\Data logger\Configuration\Storage mode
Notice!
The storage mode does not affect the outputs.
Notice!
Storage mode
=
Average
The average of the physical quantity and of other quantities assigned to the measuring channel, e.g., the measured
temperature, will be calculated.
If a storage rate < 5 s is selected,
Sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as invalid.
Tab. 18.1: Parameters for storing
display
description of the parameters
Store totalizers
values of the totalizers
Store diagnost. values
diagnostic values
Store transd. temp.
transducer temperature
Special functions\Data logger\Delete meas. values