19 Settings
19.2 Working with parameter records
PIOX R721
2021-09-01, UMPIOX_F72xV1-3EN
92
19.2
Working with parameter records
19.2.1
Introduction
Working with parameter records will make repeated measurement tasks easier and faster. Setup settings of the
transmitter are stored in the parameter records. Depending on which measuring task has to be carried out, it is possible to
load the corresponding parameter record and use it for the measurement. The transmitter can store max. 20 parameter
records.
A parameter record contains the following parameters:
• sensor parameter
• fluid parameters
• input quantities
• output quantities
• input and output parameters
19.2.2
Creation of a parameter record
The parameters have first to be entered in the following menus:
•Parameters
•Inputs
•Outputs
•Event trigger
•Data logger
•Calibration
•Communication
•Miscellaneous
Afterwards, they can be stored as parameter record.
• Select the menu item
Param. record memo.
• Press ENTER.
• Select
Save current record
.
• Press ENTER.
• Enter the name the parameter record has to be stored with.
• Press ENTER.
19.2.3
Load of a parameter record
Stored parameter records can be loaded and used for measurement.
• Select the menu item
Load param. record
.
• Press ENTER.
• Select the parameter record to be loaded.
• Press ENTER.
Notice!
No parameter records are stored in the delivery state. Parameter records have to be entered manually.
Miscellaneous\Param. record memo.
Miscellaneous\Param. record name
Miscellaneous\Param. record memo.\Load param. record