15 Data logger
PIOX R721
15.1 Configuration of the data logger
79
UMPIOX_F72xV1-3EN, 2021-09-01
15
Data logger
The transmitter has a data logger which stores the
measured values during the measurement.
The following data is stored within the data logger:
• date
• time
• measuring point number
• fluid parameters
• sensor data
• physical quantity
• unit of measurement
• measured value
Measured values transmitted via the outputs are also
stored in the data logger.
• Select the menu item
Data logger\Configuration
.
• Press ENTER.
15.1
Configuration of the data logger
Starting time
It is possible to set a starting time if it is necessary to synchronize the storing of measured values for several transmitters.
• Select the menu item
Start storing
.
• Press ENTER.
• Select the moment at which to start the storing.
Data logger
Configuration
Enable logger
Start storing
Storage rate
Ringbuffer
Storage mode
Average
Sample
Store diagnost. values
Store transd. temp.
Beep on storage
siehe Anhang A
Notice!
In order to store measured data, the data logger has
to be configured.
Data logger\Configuration\Start storing
display
description
Immediately
The storing starts immediately.
Full 5 minutes
The storing starts in the next full 5 minutes.
Full 10 minutes
The storing starts in the next full 10 minutes.
Full 15 minutes
The storing starts in the next full 15 minutes.
Full 30 minutes
The storing starts in the next full 30 minutes.
Full hour
The storing starts in the next full 60 minutes.
Event-based
The storing starts once a defined event has been occurred.