13 Outputs
13.2 Digital output
PIOX R721
2021-09-01, UMPIOX_F72xV1-3EN
74
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
Tab. 13.7: Output function test – measuring range
list item
switching function
test value
description
Status OK
(status value)
NC
Status OK
• binary output is energized
• measured value has to be low ohmic
Status error
• binary output is de-energized
• measured value has to be high ohmic
NO
Status OK
• binary output is de-energized
• measured value has to be high ohmic
Status error
• binary output is energized
• measured value has to be low ohmic
Idle state
(event value)
NC
Passive
• binary output is energized
• measured value has to be low ohmic
Active
• binary output is de-energized
• measured value has to be high ohmic
NO
Passive
• binary output is de-energized
• measured value has to be high ohmic
Active
• binary output is energized
• measured value has to be low ohmic