14 Inputs
14.2 Assignment of an input
PIOX R721
2021-09-01, UMPIOX_F72xV1-3EN
78
14.1.5
Function test of the input
The function of the installed input can now be tested.
Analog input
• Connect the signal source to the input.
• Select
Yes
to test the input signal. Select
No
to display the next menu item.
• Press ENTER.
• If the transmitter displays a value (here:
Current
), the input functions.
• Press ENTER.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the assignment of the measured value to the input signal. Select
No
to display the next menu item.
• Press ENTER.
• If the transmitter displays a value (here:
Temperature
), the input functions.
• Press ENTER.
• Select
Repeat
to repeat the test,
Finish
to display the next menu item.
• Press ENTER.
14.2
Assignment of an input
• Select the menu
Inputs
.
• Press ENTER.
• Select
Assign inputs
.
• Press ENTER.
The following physical quantities can be selected:
– Fluid temp.
– Fluid pressure
– Fluid density
• Select the physical quantity to be assigned to an input.
• Press ENTER.
• Select the input to be assigned to a physical quantity.
Only the enabled inputs are displayed in the scroll list.
• Select the list item
No linkage
if no input is to be assigned to the physical quantity.
• Press ENTER.
Inputs\...\I1 Test signal
Inputs\...\I1 Test signal\Current
Inputs\...\I1 Test mea. range
Inputs\...\I1 Test mea. range\Temperature
Assign inputs
Assign inputs\...