![Flexim FLUXUS G801 Operating Instruction Download Page 48](http://html1.mh-extra.com/html/flexim/fluxus-g801/fluxus-g801_operating-instruction_2304732048.webp)
6 Mounting
6.2 Transducers
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
48
• Repeat the steps to fix the second rail (see Fig. 6.31).
Step 5: Installation of the transducers in the mounting fixture Variofix L
• Press the transducers firmly into their clamping fixtures in the covers until they are tightly fixed. The transducer cables
show in opposite directions (see Fig. 6.32).
• Adjust the transducer distance displayed by the transmitter (see Fig. 6.33).
• Fix the transducer cables with the strain relief clamp to protect them from mechanical strain (see Fig. 6.33).
• Put coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the
transducers. The coupling foil can be fixed to the contact surface with some coupling compound.
• Put the covers with the transducers on the rails.
• Correct the transducer distance, if necessary.
Fig. 6.30: Ratchet clasp with tension strap
1 – sense of rotation
2 – edge
3 – lever
4 – clasp screw with slot
Fig. 6.31: Pipe with 2 rails
Notice!
The arrows on the transducers and the covers have to point in the same direction.
Fig. 6.32: Installation of the transducers in the covers
1 – cover
2 – transducer clamping fixture
1
2
4
3
4
1
2
1
2