16 Advanced functions
16.9 Modification of the limit for the inner pipe diameter
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
138
• Activate or deactivate the auto-snap function.
• Press ENTER.
• Activate or deactivate the snap ringbuffer.
• Press ENTER.
Delete snaps
• Select
Clear Snaps
.
• Press ENTER.
Read snaps
• Select
Snaps ->Rs232
.
• Press ENTER.
Activation of the snap function
• In order to activate the snap function, press
until
DSPSNAP/VOLTAGE
is displayed in the upper line.
• Press ENTER.
16.9
Modification of the limit for the inner pipe diameter
It is possible to modify the lower limit of the inner pipe diameter for a given transducer type.
• Enter the HotCode
071001
.
• Enter the lower limit of the inner pipe diameter of the displayed transducer type.
• Press ENTER to select the next transducer type.
Special Funct.\...\DSP-SignalSnap\AutoSnap
Special Funct.\...\DSP-SignalSnap\Snap ringbuffer
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap\Clear Snaps
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap\Snaps ->Rs232
Notice!
If a transducer is used below its recommended inner pipe diameter, a measurement might be impossible.
DNmin Q-Sensor
15
mm