16 Advanced functions
FLUXUS G801
16.8 Diagnosis with the help of the snap function
137
UMFLUXUS_G801V1-0EN, 2020-06-26
16.7.4
Measurement with calculation channels
• Select the program branch
Measuring
.
• Press ENTER.
• Activate the necessary channels. Calculation channels are activated or deactivated the same way as measuring channels.
• Press ENTER.
If a measuring channel that is needed for an activated calculation channel has not been activated, a warning will be
displayed.
• Press ENTER.
Position the transducers for all activated measuring channels. The measurement will be started automatically.
If a calculation channel is activated, the HumanMux mode will be selected at the beginning of the measurement and the
values of the calculation channel will be displayed.
If the AutoMux mode is selected, the measured values of the measuring channels will be displayed alternately (but not the
measured values of the calculation channels).
• Press
to display the calculation function.
• Press
to display the measured values of the different channels.
16.8
Diagnosis with the help of the snap function
By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of
measuring results or for diagnostic purposes.
• Select the menu item
Signal snap
.
• Press ENTER.
Snap memory settings
• Select
on
to activate the snap function. Select
off
to deactivate the snap function.
• Press ENTER.
• Select
Install Snap
.
• Press ENTER.
• Enter the number of snap memory storage spaces.
• Press ENTER.
Measuring\CHANN: A B Y Z
Measuring\...\WARNING! CHANNEL B:INACTIV!
Special Funct.\SYSTEM settings\Signal snap
Special Funct.\SYSTEM settings\Signal snap\DSP-SignalSnap
Special Funct.\...\DSP-SignalSnap\Install Snap
Special Funct.\...\DSP-SignalSnap\Install Snap\Snap-Memory
Y:FLOW VELOCITY
53.41
m/s