6 Mounting
6.2 Transducers
FLUXUS G801
2020-06-26, UMFLUXUS_G801V1-0EN
46
• Select the installation instruction of the supplied clasp:
Band clamp clasp
• Insert the tension strap into its clamp (2) (see Fig. 6.26).
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.27).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of the tension strap clamp (2), but not too firmly in order not to damage the tension strap (see Fig. 6.26).
Quick release clasp
• Insert the tension strap into its clamp (2) and the metal spring (see Fig. 6.26 and Fig. 6.28).
• Place the tension strap around the pipe and insert it into the clasp.
• Position the metal spring opposite the tension strap clamp (2).
• Tighten the tension strap.
• Tighten the clasp screw.
• Tighten the nut of the tension strap clamp (2), but not too firmly in order not to damage the tension strap (see Fig. 6.26).
Fig. 6.25: Rail, fixed on one side to the pipe
1 – tension strap clamp
2 – tension strap clamp
3 – nut
Fig. 6.26: Rail on the pipe
1 – tension strap clamp
2 – tension strap clamp
3 – metal spring
4 – nut
5 – clasp
2
1
3
1
3
4 2
5
1