9 Measurement
FLUXUS F502TE
9.4 Display of measured values
89
UMFLUXUS_F502TEV1-5EN, 2020-05-29
9.4.3
Transducer distance
The transducer distance can be displayed during the measurement by scrolling with
.
The optimum transducer distance (here: 51.2 mm) will be displayed in parentheses, followed by the entered transducer
distance (here: 50.8 mm).
The optimum transducer distance might change during the measurement (e.g., due to temperature fluctuations).
A deviation from the optimum transducer distance (here: 0.4 mm) will be compensated internally.
Tab. 9.5:
Description of the status line
value
explanation
S
signal amplitude
0
…
9
< 5 %
…
≥ 90 %
Q
signal quality
0
…
9
< 5 %
…
≥ 90 %
c
sound speed
comparison of the measured and the expected sound speed of the fluid
The expected sound speed is calculated from the fluid parameters.
√
OK, is equal to the expected value
↑
> 20 % of the expected value
↓
< 20 % of the expected value
?
unknown, cannot be measured
R
flow profile
information about the flow profile based on the Reynolds number
T
fully turbulent flow profile
L
fully laminar flow profile
↕
transition range between laminar and turbulent flow
?
unknown, cannot be calculated
F
flow velocity
comparison of the measured flow velocity with the flow limits of the system
√
OK, the flow velocity is not within the critical range
↑
the flow velocity is higher than the current limit
↓
the flow velocity is lower than the current cut-off flow
0
the flow velocity is within the limit range of the measuring method
?
unknown, cannot be measured
Fig. 9.3:
Display of the transducer distance
Notice!
Never change the transducer distance during the measurement.
L=(51.2) 50.8 mm
54.5
m3/h