7 Connection
7.4 Temperature probe
FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
64
Connection to the transmitter
• Remove the blind plug for the connection of the inline temperature probes (see Fig. 7.13).
• Prepare the cables (see Fig. 7.20).
• Open the cable gland.
• Push the cables of the inline temperature probes through the cap nut, the compression part and the basic part.
• Insert the cables into the housing.
• Screw the sealing ring side of the basic part into the transmitter housing.
• Brush the shields of the inline temperature probe cables back over the compression part. The shields and the basic part
of the cable gland must have direct contact.
• Fix the cable gland by inserting the compression part into the basic part and by screwing the cap nut onto the basic part.
• Connect the inline temperature probes to the terminals of the transmitter (see Fig. 7.13 and Tab. 7.14).
Fig. 7.20: Cable gland and preparation
1 – compression part
2 – shield
3 – cap nut
4 – basic part
Tab. 7.14: Terminal assignment (transmitter)
terminal
extension cable
TF1, TX1
red
TF2, TX2
gray
TF3, TX3
blue
TF4, TX4
white
4
3
1
80 mm
7 mm
10 mm
2