Flex-Pro Norprene A3 Operating Manual Download Page 18

Page 18

9.3

0 - 10 VDC Input (Volt DC)

Used to remotely control pump with an incoming 0-10 VDC signal.

Default settings:

0 VDC = 0% motor speed
10 VDC = 100% motor speed

Press SELECT RUN MODE button until 

0 - 10 VDC Input i

s displayed in the top 

line of the display. 

To configure the pump, navigate to 

0 - 10 VDC Input

 menu by using short-cut 

method described at beginning of section, or by pressing MENU button, then 
selecting Input Setup, Input Modes, and then 

0 - 10 VDC Input

.

Press UP or DOWN arrow to scroll through 0 - 9 on selected digit.

Press RIGHT arrow to scroll over to next digit to right. If you pass your desired 
digit, you can easily scroll back by continuously pressing RIGHT button.

Press ENTER to save changes.

Continue this process until all four screens have been configured.

If you used short-cut to enter 0 - 10 VDC Input setup, then just press and hold 
Select Run Mode button until Run Mode screen is displayed.

If you used Menu button to navigate to 0 - 10 VDC Input setup, you must 
navigate back out of menu structure. To do this you must select <-Esc at bottom 
of every screen menu until you see Run Mode screen displayed.

- 0-10 VDC Input

% Motor @ Min Signal:

Input % Motor Speed at
Minimum Volt DC Signal

00.0%

0

- 0-10 VDC Input

Min VDC Signal:

Input Min. Volt DC signal

0.00

0

- 0-10 VDC Input

% Motor @ Max Signal:

Input % Motor Speed at
Max VDC Signal

00.0%

1

- 0-10 VDC Input

Max VDC Signal:

Input Max. Volt DC signal

0.00

1

Press ENTER to set.

Press ENTER to set.

In 0 - 10 VDC mode, you can view current 
incoming signal by pressing RIGHT arrow (see 
screen shot). RIGHT arrow is a convenient 
way to scroll through multiple read-only 
screens during normal operating mode of 
pump.

Tip!

0-10 VDC Input

VDC 

Displays current incoming signal

12.5

- Screen Shot -

Å

É

Ä

È

Displays current rotor direction

Example 1

0

2

4

6

8

10

0

25

50

75

100

Pump

Motor Speed (%)

VDC Input

from external source

0 VDC = 0% Pump Output

10 VDC = 100% Pump Output

Example 2

0

25

50

75

100

0 VDC = 100% Pump Output

10 VDC = 0% Pump Output

Example 3

0

25

50

75

100

0 VDC = 20% Pump Output

8 VDC = 45.3% Pump Output

0%

Output

Pump

Motor Speed (%)

Pump

Motor Speed (%)

100%

Output

100%

Output

0%

Output

45.3%

Output

0

2

4

6

8

10

VDC Input

from external source

0

2

4

6

8

10

VDC Input

from external source

20%

Output

0%

Output

100%

Output

100%

Output

0%

Output

50%

Output

20%

Output

Flex-Pro

Summary of Contents for Norprene A3

Page 1: ...e co id or info spe co id URL www spe co id SERIES A3V and A4V Operating Manual ProSeries by Blue White Ind TM Patents 7 001 153 7 284 964 4 496 295 and other patents pending FLEX PRO Peristaltic Mete...

Page 2: ...onal Dosing 24 10 0 Contact Closure Input 25 11 0 Set FVS flow verification system 26 12 0 Output Setup Signal Relay Contact Closure 27 12 1 Signal Output 28 13 0 Pump Tube Timer 29 14 0 TFD tube fail...

Page 3: ...C 130 54 130 54 F C 130 54 130 54 130 54 130 54 F C 185 85 185 85 185 85 185 85 185 85 Norprene A3 Tube Pumps Meets FDA criteria for food Excellent chemical resistance Extra long life at low pressures...

Page 4: ...15 1 4 D 20 6W x 27 3H x 38 9D cm A4 models 12 1 8 W x 14 1 4 H x 18 5 8 D 30 8W x 36 1H x 47 3D cm Approximate shipping wt A3 models 33 lb 15 0 Kg A4 models 56 lb 25 4 Kg Wetted components Non Wette...

Page 5: ...15V 1A contact closures assignable to monitor up to 17 different pump functions including TFD FVS remote local control setting motor on fault current operating mode and others Two CNC precision machin...

Page 6: ...nly n Raise metering pump 4 1 2 inches 11 43 cm off ground or a surface n Made out of tough Stainless Steel n Provides a stable mounting surface Stainless Steel extended brackets allow pump to be secu...

Page 7: ...buildup when injecting chemicals that calcify such as sodium hypochlorite Duckbill may add additional back pressure to pump up to 7 psi 3 8 OD x 1 4 ID Tubing and Tri Clamp connection Models Proper ey...

Page 8: ...and power cables together Power Interruption This pump has an auto restart feature which will restore pump to operating state it was in when power was lost Note When in doubt regarding your electrical...

Page 9: ...CE ACTIVE LOOP OPEN COLLECTOR 0 1000 Hz 50 DUTY CYCLE 6 AMP MAX AT 250 VAC 5 AMP MAX AT 30 VOLT DC 1 AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC 1 AMP MAX AT 125 VOLT AC 0 8 AMP MAX AT 30 VOLT DC...

Page 10: ...a runtime function and is also used when configuring your pump Runtime Function While pump is operating in any Run Mode you can view useful information regarding your current Run Mode by pressing RIG...

Page 11: ...until STOP or START button is pressed STOP button will stop pump and display will read OFF START button will stop prime feature and jump back into previous operating state To REVERSE ROTATION of rotor...

Page 12: ...ist of all available options Scroll bar example While making a selection where only one choice is allowed you will see a radio button Radio button example In a screen where you are making changes you...

Page 13: ...lay Language Esc Reset to Defaults Set Password Display Rate Reset to Factory Defaults Reset to Factory Defaults Reset to Factory Defaults Are you sure Yes No Liter Hr Gal Hr US Gal day US Done Displa...

Page 14: ...Display Rate Press MENU button to enter menu structure for setting up pump Select Configuration and Press ENTER button Select Display Rate and Press ENTER button Select your desired Display Rate unit...

Page 15: ...fault setting is DISABLE Set FVS Flow Verification System Set Flow Verification time delay Use this feature if you are using a Blue White flow verification sensor to monitor flow output Default settin...

Page 16: ...nu structure To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed Tip This feature can be combined with Contact Input feature to allow for remote Start...

Page 17: ...t Esc at bottom of every screen menu until you see Run Mode screen displayed 4 20 mA Input Motor Min Signal Input Motor Speed at Minimum mA Signal 00 0 0 4 20 mA Input Min mA Signal Input Min mA signa...

Page 18: ...ture To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed 0 10 VDC Input Motor Min Signal Input Motor Speed at Minimum Volt DC Signal 00 0 0 0 10 VDC I...

Page 19: ...ency Input setup you must navigate back out of menu structure To do this you must select Esc at bottom of every screen menu until you see Run Mode screen displayed Frequency Input Motor Min Signal Inp...

Page 20: ...hold Select Run Mode button until Run Mode screen is displayed If you used Menu button to navigate to Pulse Batch setup you must navigate back out of menu structure To do this you must select Esc at...

Page 21: ...setup then just press and hold Select Run Mode button until Run Mode screen is displayed If you used Menu button to navigate to Cycle Adjustment setup you must navigate back out of menu structure To...

Page 22: ...arrow to scroll through 0 9 on selected digit Press RIGHT arrow to scroll over to next digit to right If you pass your desired digit you can easily scroll back by continuously pressing RIGHT button P...

Page 23: ...desired digit you can easily scroll back by continuously pressing RIGHT button Press ENTER to save changes Continue this process until all four screens have been configured If you used short cut to e...

Page 24: ...easily scroll back by continuously pressing RIGHT button Press ENTER to save the changes Continue this process until all four screens have been configured If you used short cut to enter Proportional D...

Page 25: ...croll down to Done selection Then press ENTER To navigate back out of menus select Esc and press the ENTER button at bottom of every screen menu until you see Run Mode screen displayed IMPORTANT If Co...

Page 26: ...d during stopper operation such as with chlorine an alarm delay time value from 1 255 seconds must be programmed An alarm delay value of 000 seconds disables FVS system If FVS alarm occurs pump will s...

Page 27: ...ontact 1 Contact 2 Contact 3 Esc Relay Output 4 20 mA Output Signal min Motor Input mA Signal Min Motor speed 4 20 mA Output Minimum Motor Input Min Motor for mA Signal Output 4 20 mA Output Signal Ma...

Page 28: ...mA 100 Pump Output 20 mA Signal Output Frequency Output Disable Done 4 20 mA Output Default 4 20 mA output settings shown Signal Output Disable Done 4 20 mA Output Default Frequency Hz output settings...

Page 29: ...14 0 TFD Tube Failure Detection Flex Pro is equipped with a Tube Failure Detection System which is designed to stop pump and provide an output alarm see Output menu in the event pump tube should rupt...

Page 30: ...Stop pump by pressing STOP button Thoroughly clean Pump Head and Rotor Rotor can be removed by pulling straight out After cleaning process push Rotor back on shaft See drawing above for proper assembl...

Page 31: ...to slide discharge fitting into pump head tongue and groove Once discharge fitting is secured in pump head stop pump by pressing STOP button Replace pump head cover Pump will ask you if you d like to...

Page 32: ...Page 32 This Page Blank Flex Pro...

Page 33: ...Page 33 This Page Blank Flex Pro...

Page 34: ...Page 34 This Page Blank Flex Pro...

Page 35: ...to allow back flow If your desire is to simply extend tube life Typically tubing fails on outlet side pressure side of tube assembly in pump head Failure to install checkvalves in their proper flow d...

Page 36: ...mical from pump Determine which parts are being attacked and replace them with parts that have been manufactured using more suitable materials Manufacturer does not assume responsibility for damage to...

Page 37: ...ller assemblies rotors in this group are same size and interchangeable Tubing in this group are interchangeable Norprene Group All roller assemblies rotors in this group are same size and interchangea...

Page 38: ...Tubing ID Example Tube Assembly reorder number with NH printing A3 SNH T A 3 S N H T Pump Series Tubing Material N Norprene T Norprene Chemical G Tygothane Tubing ID D 075 E 125 F 155 G 187 H 250 J 3...

Page 39: ...tested with water only for pressure and performance Installers and operators of these devices must be well informed and aware of the precautions to be taken when injecting various chemicals especiall...

Reviews: