SYMBOLS
Class II unit
Type-B unit
Important: check the operating
instructions
Alternate current
Switch oFF
Switch oN
Do not use the unit while bathing
or showering
manufacturer
N
ON
-AL
LERGIC MATE
RIA
L
BIO
COMPATIB
LE
In compliance with European standard
EN 10993-1 Biological evaluation of medical
devices and 93/42/EEC Medical Devices Directive
TECHNICAL SPECIFICATIONS
Mod. FLAEM JUNIOR F2000
Voltage:
230V~ 50Hz 210VA
115V~ 60Hz 230VA 220V~ 60Hz 230VA 100V~ 50/60Hz 230VA
In conformity Dir. 93/42/EEC
with:
(EN 60 601-1)
Approvals:
Description
Code
- RF6 PLUS nebuliser kit composed of:
nebuliser, mouthpiece with valve and nasal prong
ACo368P
- 2 m connection tube ACo225P
- Mask Set composed of: adult mask, pediatric mask and elastic band ACo462P
- Spare air filter kit (2 pcs) ACo164P
AVAILABLE SPARE PARTS
- After sanitization shake the accessories vigorously and place them on a paper towel, or alternatively,
dry them with a jet of hot air (e.g. using a hair dryer).
- Dispose of the solution as recommended by the disinfectant manufacturer.
STERILIZATION
The accessories that can be sterilized are (
C5.1-C5.2-C5.3-C5.4-C3-C4)
The sterilization procedure described in this paragraph is effective only if all the prescriptions herein
are met, all the components involved are previously sanitized and the procedure itself is validated
according to ISo 17665-1 rule.
Unit:
Steam sterilizer with fractionated vacuum and overpressure according to EN 13060 rule.
Procedure:
Package each component to be treated in sterile barrier system or packaging complying
with EN 11607 rule. Place the packed components in the steam sterilizer.
Run the sterilization cycle according to the instructions manual of the device, selecting a temperature
of 134° C and a duration of 10 minutes
Storage:
Store sterilized components as per instructions of the sterile barrier package or system chosen.
AIR FILTER
The device has a suction filter (A3) that should be replaced when it is dirty
or changes color. Do not wash or reuse the same filter. Regular replacement
of the filter is required to help and ensure correct performance of the
compressor. Check the filter on a regular basis. Contact your local dealer or
authorized service center for spare filters.
Use original FLAEM accessories only.
9
Max Pressure: 3.5 ± 0.5 bar
Compressor air output: 14 LPM approx
Noise level (at 1 m): 57 dB (A) approx
operation: Continuous use
Dimensions 20(W)x18(D)x11(H) cm
Weight 1.8 Kg
Summary of Contents for HospiNeb FLAEM JUNIOR F2000
Page 2: ......
Page 34: ...32 1 A7 1 5 1 2 5 3 5 1 3 4 1 5 6 RF6 Plus RF6 Plus C5 4 1 5 ml RF6 Plus C5 4 2 C4 1 2...
Page 38: ...36 1 A 2 A 3 A 4 A 5 A 6 A 7 A...
Page 39: ...37...
Page 40: ...38...
Page 41: ...39 PHTHALATES BPA FREE N O N ALLERGIC MATE R I A L B IOCOMPATIBL E...
Page 42: ...40 C1 C2 C3 C4...
Page 43: ......