Chapter 4 | Calibration
4.1.3 pH Calibration Example
The following is an example of a three point pH calibration
using auto-recognized pH 4.01, 7.00 and 10.01 pH
buffers. The pH buffers can be read in any order. For this
example, pH buffers are read lowest to highest.
Press the “Cal” key to start the pH calibration. The
active Calibration Buffer Set is shown.
Rinse the pH electrode and ATC probe and place
into the pH 4.01 buffer.
Wait for the pH value to stabilize.
Once the reading is stable, press the “Enter” key.
Rinse the pH electrode and ATC probe and place
into the pH 7.00 buffer.
Wait for the pH value to stabilize.
Once the reading is stable, press the “Enter” key.
Rinse the pH electrode and ATC probe and place
into the pH 10.01 buffer.
Wait for the pH value to stabilize.
Once the reading is stable, press the “Enter” key.