7.1 Meter Specifications
pH
Range
-2.00 to 18.00 pH
Resolution
0.1, 0.01 pH
Relative Accuracy ±0.01 pH ±1 LSD
Calibration Points 1 to 5 points
Calibration
Method
Automatic buffer recognition with
manual buffer entry option
Calibration pH
Buffer Sets
USA: 2.00, 4.01, 7.00, 10.01, 12.00
NIST: 1.68, 4.01, 6.87, 9.18, 12.46
DIN: 1.09, 3.06, 4.65, 6.79, 9.23, 12.75
FSCI: 1.00, 3.00, 6.00, 8.00, 10.00, 13.00
Slope Display
Yes
mV / RmV
Range
±2000 mV
Resolution
0.1 mV
Relative Accuracy
±0.2 mV or ±0.05 % of the reading,
whichever is greater
Calibration Points 1 point
Offset Adjustment Up to ±250 mV
Temperature
Range
-5.0 to 105.0 ºC, 23.0 to 221.0 ºF
Resolution
0.1 ºC, 0.1 ºF
Relative Accuracy ±0.3 ºC, ±0.5 ºF
Calibration Points 1 point
Offset Adjustment Up to ±5 ºC, up to ±9 ºF
Features
Display
5” backlit LCD
Time and Date
Yes
Read Types
Continuous, Auto-Read, Timed
Timed Interval
5 seconds to 60 minutes
Hold Function
Yes
Data Log
500 data sets with time and date
Calibration Log
Active pH, RmV, temperature
Calibration Alarm Yes
Sleep Mode
Yes
Memory
Non-volatile
Input
BNC, ATC
Data Output
Computer or printer
Data Output
Format
CSV or print
Warranty
3 years
Certifications
CE, TUV 3-1, FCC Class A
Enclosure
IP-54
Power
100-240 VAC, 50-60Hz, 9 DC
adapter, 1.3A
Dimensions
(L x W x H)
198.3 mm x 155.2 mm x 61.1 mm,
7.81” x 6.11” x 2.41”
Weight
700 grams, 1.54 lbs.
Environmental Conditions
Environmental
Conditions
Indoors
Altitude
Up to 2,000 Meters
Operating
Temperature
5°C to 45°C
Operating Relative
Humidity
5 to 85%, non-condensing
Storage
Temperature
-20°C to 60°C
Storage Relative
Humidity
5 to 85%, non-condensing
Mains Fluctuation ±10% of range (100-240VAC)
Installation
Category
II
Pollution Degree
2
Protection Class
III
Summary of Contents for accumet AB315