accumet AB315 pH/mV Meter 19
The minutes value will flash. Press the
or
key
to adjust the minutes.
a.
Press the
key to increase by one minute.
Press and hold the
key to quickly increase.
b.
Press the
key to decrease by one minute.
Press and hold the
key to quickly decrease.
Press the “Enter” key to save the change.
3.2.15 Audio Mode
Set the audible beep on or off. The audible beep is used
each time a meter key is pressed, when a measurement
stabilizes and when an alarm is triggered.
Press the
or
key to scroll through the Setup
Menu list until the Audio Mode item is shown.
Press the
or
key to set the audio mode.
Press the “Enter” key to save the change.
3.2.16 Sleep Mode
Set the sleep mode on or off. When the meter sleep mode
is on, the meter will enter sleep mode when no keys are
pressed for 20 minutes. Once the meter is in sleep mode,
press the “Power” key to resume using the meter. Make
sure sleep mode is off when taking timed measurements.
Press the
or
key to scroll through the Setup
Menu list until the Sleep Mode item is shown.
Press the
or
key to set the sleep mode.
Press the “Enter” key to save the change.
3.2.17 Clear Data
Erase all data logs or erase all calibration logs
Press the
or
key to scroll through the Setup
Menu list until the Clear Data item is shown.
Press the
or
key to select the data log (dAtA)
or calibration log (CAL) to be cleared.