2.4.4.1 Take the dispenser and connection tubing bundle out on the bench.
Connect one water
tubing on the bundle
as marked to the
system outlet (To
POU) and inlet (In
from system) to the
dispenser. Tubing is
marked with
corresponding tags
(IN and TO POU) .
Connect the other
marked tubing to the
inlet (FROM POU) on
the system and the
other end (Out to
system) to the
Dispenser. Tubing
is marked with
corresponding tags
(OUT and FROM
POU).
Connect the 9 Pin
cable in the bundle
between the
dispenser and the
system. Tighten
the nuts.
Note:
Do not install final filter
at this stage. It will be installed after system
start-up and degassing.
End of System Installation
Proceed to "System Start Up and Operation"