c) Remove the system back cover
d) Find the UV lamp chamber (see
picture). Remove the UV lamp cover
to expose the UV lamp.
e) Unplug the UV lamp from its power
cord. Carefully remove the old UV
lamp out.
f) Wear gloves included in the new UV
lamp package. Avoid direct skin
contact with the quartz glass of the
UV lamp. Carefully insert the new UV
lamp into its chamber. When about
2/3 of the lamp is inserted, hold the
UV lamp and connect it to the ballast
cable (4-pin connector) as shown in
the picture, and then gently insert the
UV lamp
cover
UV lamp
chamber
4-pin connector
UV lamp
UV lamp