background image

 

 

 

Operations Manual 

Electrochemistry Meters 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

68X002128 | Revision 01 | September 2021

 

accumet AB330 Conductivity  
Meter User Manual 

 

Summary of Contents for Fisherbrand accumet AB330

Page 1: ...Operations Manual Electrochemistry Meters 68X002128 Revision 01 September 2021 accumet AB330 Conductivity Meter User Manual...

Page 2: ......

Page 3: ...3 1 Setup Menu Scrolling 8 2 3 2 Setup Menu Value Changes 8 2 3 3 Mode Shortcut Key 8 2 3 4 Date and Time View Shortcut Key 8 2 4 Display Overview 9 2 5 Meter Connections 10 2 6 Electrode Information...

Page 4: ...ue 19 3 2 16 Set Time Value 20 3 2 17 Audio Mode 20 3 2 18 Sleep Mode 20 3 2 19 Clear Data 21 3 2 20 Factory Reset 21 Chapter 4 22 Calibration 22 4 1 Conductivity Calibration 22 4 1 1 Conductivity Cal...

Page 5: ...30 Meter Info 30 7 1 Meter Specifications 30 7 2 Meter Dimensions 31 7 3 Warranty 31 7 3 1 Return of Items 31 Chapter 8 32 Regulatory Compliance 32 8 1 European Union 32 8 2 Product Safety 32 8 3 Elec...

Page 6: ...Note This symbol in the context of a Note indicates to follow the manufacturer specified instructions notes and requirements set out in the instruction manuals CAUTION Before using or maintaining thi...

Page 7: ...ith US Japan Euro UK Singapore Australia New Zealand and China wall plug plates is included with the meter This universal power adapter is specifically for use with this meter Use of other power adapt...

Page 8: ...ate and time setting Press to save a setting in the Setup Menu Press to accept a calibration point in the Calibration mode Press to power on the meter When the meter is on press to turn the backlight...

Page 9: ...ter is in the Measure mode Memory Icon Indicates meter is in the Setup Menu and viewing the data log or calibration log Log Saved Icon Indicates meter is actively saving data to the data log or calibr...

Page 10: ...purchase electrode separately or kitted with a conductivity electrode Refer to the electrode instructions for specific care and maintenance procedures The following are general recommendations for el...

Page 11: ...saved data logs and calibration logs to the selected export device 4 Temperature Calibration or Manual Temperature Input When an ATC probe is connected use the Temp Calibration menu to perform a temp...

Page 12: ...pressed the View Logs setup menu item is always shown first Press the or key to scroll through the data and calibration log options Press key Press key Press key Press key Press key Press the Enter ke...

Page 13: ...he Enter key to view the display log For the calibration log press the or key to change the view from the date time screen to the calibration data screen Press or key For the calibration log press the...

Page 14: ...d export device 3 2 4 Temperature Calibration or Manual Temperature Input 3 2 4a Temperature Calibration When an ATC probe is connected use the Temperature Calibration menu to perform a one point temp...

Page 15: ...etup Menu list until the Measure Mode item is shown Press the or key to set the measure mode Press the Enter key to save the change 3 2 6 Nominal Cell Constant Set the nominal value for the cell const...

Page 16: ...inear temperature compensation value enter a value of 0 0 to 10 0 C Press the or key to scroll through the Setup Menu list until the Temperature Compensation Coefficient item is shown Press the or key...

Page 17: ...Export key to save a measurement to the data log and export Auto Read In the Auto Read measure mode press the Read key to start a measurement When the measurement is stable the checkmark stable icon i...

Page 18: ...time interval Press the or key to scroll through the Setup Menu list until the Calibration Due Alarm item is shown Press the or key to adjust the alarm value a Press the key to increase the alarm valu...

Page 19: ...he month value will flash Press the or key to adjust the month a Press the key to increase by one month Press and hold the key to quickly increase b Press the key to decrease by one month Press and ho...

Page 20: ...the key to decrease by one minute Press and hold the key to quickly decrease Press the Enter key to save the change 3 2 17 Audio Mode Set the audible beep on or off The audible beep is used each time...

Page 21: ...onfirm when the selected log is fully erased The meter will return to the measurement screen 3 2 20 Factory Reset Erase all data logs calibration logs and settings and return the meter to its factory...

Page 22: ...ll determine how the conductivity calibration standards are displayed during the calibration If temperature compensation is set to Off Once the calibration reading is stable the meter will display the...

Page 23: ...alue is shown press the Enter key to accept the conductivity standard value and proceed to the next calibration point To calibrate with another conductivity standard repeat steps 3 5 or to save and en...

Page 24: ...ept the TDS standard value To calibrate with another TDS standard repeat steps 3 6 or to save and end the calibration press the Cal key The meter will display the calculated calibration factor value a...

Page 25: ...T traceable thermometer Press the Menu key to enter the Setup Menu Press the or key to scroll through the Setup Menu list until the Temp Calibration item is shown The temperature value will blink indi...

Page 26: ...eter is in the measurement mode 5 2 Conductivity Salinity and TDS Measurements Rinse the electrode with distilled water or appropriate solution and blot gently with a lint free tissue to remove excess...

Page 27: ...saved with its associated date and time stamp To view the time stamp with seconds included the data log must be exported to a computer or printer The conductivity calibration log and TDS calibration l...

Page 28: ...mode MODE COND or SALT or TDS Remote Command GETLOG CR Prints all logged measurement data If no data is logged returns to receive next command Output format is based on the Log Export Type setting Rem...

Page 29: ...sing plates and temperature sensor submerged below the top of the sample solution o Verify cables do not have any damage o Perform a factor reset on the meter 6 2 Measurement Troubleshooting Reading i...

Page 30: ...p to 9 F Features Display 5 backlit LCD Time and Date Yes Read Types Continuous Auto Read Timed Timed Interval 5 seconds to 60 minutes Hold Function Yes Data Log 500 data sets with time and date Calib...

Page 31: ...epaired on a charged basis The warranty on your meter shall not apply to defects resulting from Improper or inadequate maintenance by customer Unauthorized modification or misuse Operation outside of...

Page 32: ...ironment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communicat...

Page 33: ...der entsorgt werden Mehr Informationen ber die Einhaltung dieser Anweisungen durch Thermo Scientific dieVerwerter und Hinweise die Ihnen n tzlich sein k nnen die Thermo Fisher Scientific Produkte zu i...

Page 34: ...____________________ __________________________________________________ __________________________________________________ __________________________________________________ __________________________...

Page 35: ..._______________ __________________________________________________ __________________________________________________ __________________________________________________ _______________________________...

Page 36: ...40V 13 637 690 accumet RS232 cable for thermal compact printer 13 637 681 accumet paper for compact printer 2 pack 13 637 691 Find out more at fishersci com Distributed by Fisher Scientific Contact us...

Reviews: