19
Also, before using USB3 connection, the C-RED 3 USB3 drivers must be installed on the
PC.
The drivers are included in the USB SDK library. By default, this library is installed with the GUI demo
software. Please simply install the demo software and the drivers will be installed.
Note: To have the full USB3 bandwidth, it is needed for some PC to disable all the energy saving settings
(cf. C-RED 2-3 TS2 disable power saving for USB3). Also, to get the full USB3 bandwidth, it is usually
required to use native USB3 port.
FliVision Graphical User Interface
The Graphical User Interface (GUI) Flivision is provided in the USB key supplied with the C-RED 3
camera, or available on demand at
It is a dedicated interface developed by First Light Imaging which allows to monitor and control all the
parameters of the camera. This software has its own manual.
Powering up/down
Power ON:
When the power FISCHER CONNECTORS is connected to the camera, and the power supply to the line
plug, the camera is ON.
Power OFF:
Please use the CLI command “
shutdown
” from a simple terminal before turning off the camera. First
unplug the power supply from the line plug, then unplug the FISCHER CONNECTORS cable from the
camera.
The shutdown command is recommended (specially to store the latest logs), however, the direct switch
off does not damage it.