36
FPS_Tint_in timing diagrams
Standard polarity / externally controlled integration time
When external signal polarity is set to standard, and integration duration is controlled externally, the
falling edge triggers the start of integration, the rising edge triggers the frame readout and the time
between the falling edge and the rising edge is the integration time.
Figure 23:
Standard polarity / externally controlled integration time
Inverted polarity / externally controlled integration time
When external signal polarity is set to inverted, and integration duration is controlled externally, the
rising edge triggers the start of integration, the falling edge triggers the frame readout and the time
between the rising edge and the falling edge is the integration time.
Figure 24:
Inverted polarity / externally controlled integration time
Non-integrating time.
Min value: 3µs
Start of
integration
FPS_Tint_in
Start of
reading
Integration time
FSYNC_out
Start of
integration
FPS_Tint_in
Start of
reading
Integration time
FSYNC_out
Non-integrating time.
Min value: 3µs