Confirm by pressing
, then
and
to select another day
4.3.3.5
Setting the user password
This menu gives the possibility to set a user password to protect DMLP Digital Monitor switching on/off, setting of
high and low temperature limit, date and time changing and accessing to MODE menu.
DMLP Digital Monitor is factory set with password = 00.
Press the button
, then use
or
till the following string
.
To set the password, press
and select a number from 0 to 255 using
and
buttons. Then, confirm
.
To change the password, press the button
. The actual password is now required (if any password has been set,
the display won’t ask it).
Use
or
to enter the password and press
. If the password is correct,
it’s possible to
modify the old
one and confirm it by pressing
. On the contrary, if password is incorrect, the display will show PASSWORD
ERROR and the controller will automatically go to the main menu.
If password is set again to (00), switching on/off, limits and date/time modification
won’t be protected anymore by
password.
BEWARE: IF THE USER PASSWORD IS LOST, THERE IS NO POSSIBILITY TO RETRIEVE
IT.
4.3.3.6
Setting the date and time
Press the button
, then use
or
till the following string
.
Confirm
and the display will blink the actual time. Modify the values using the button
or
and
confirm the new time with
. Continue until all the values are set.
BEWARE:
We suggest to set (if necessary) date and time only when the equipment is switched on for the
first time, and avoid modifying them later. Changing DMLP Digital Monitor date and time means an
immediate erasing of the recorded data, i.e. those referred to the latest 45 days recorded. Data won’t be lost
anyway, since it will always be stored and available in the SD Card. If you wish to modify date and time, the DMLP
Digital Monitor records the date of the last MEMORY RESET and it is possible to access quickly this function through
the TEST function.
To grant a proper recording, the DMLP Digital Monitor shows in the display the following blinking alarm:
H: 00:00 m 00:00
(clock battery low voltage).
DOOR 05/07/06
PASSWORD m5
CLOCK SET m6
Summary of Contents for 130