Section INDUSTRIAL . ISOLATED SIGNAL CONVERTERS
17
13. Configuration menu (cont.)
Key ‘UP’
(‘force’ menu)
The key ‘
UP
’ (
5
) at the front of the instrument gives access to a
configurable list of functions (see section 12.4).
At the ‘
Key UP (‘force’ menu)
’ (
K.uP
) menu select which functions will
be available when pressing the front key ‘
UP
’ (
5
). Select ‘
on
’ to activate
the desired functions.
•
configure ‘
Force Low
’ (
F.Lo
) to ‘
on
’ to activate the ‘
Force low
’ function
menu entry.
•
configure ‘
Force High
’ (
F.hI
) to ‘
on
’ to activate the ‘
Force high
’ function
menu entry.
•
configure ‘
Force Set
’ (
F.SEt
) to ‘
on
’ to activate the ‘
Force set
’ function
menu entry.
•
configure ‘
Tare
’ (
tArE
) to ‘
on
’ to activate the ‘
Tare
’ function menu
entry.
The functions configured to ‘
on
’ are available at the ‘
force’
menu. See
section 12.4 for a description on each function and how to operate them.
13.7 Key ‘UP’ (‘force’ menu)
Force low
Force high
Force set
Tare
The key ‘
LE
’ (
3
) at the front of the instrument gives access to a
configurable set of information messages.
At the ‘
Key LE (messages function)
’ (
K.LE
) menu, select the informations
to be displayed when the front key ‘
LE
’ (
3
) is pressed (see section 12.5).
Select ‘
on
’ to activate each information.
•
configure ‘
mV at terminals
’ (
MEAS
) to ‘
on
’ to see the value in
millivolts at terminals (for example : ‘
MEAS mV 13.82
’).
•
configure ‘
Tare value
’ (
tArE
) to ‘
on
’ to see the actual value of the ‘
tare
’
parameter. (for example : ‘
tArE mV 1.27
’) This value is expressed in
corrected millivolts (mV’) (see section 10.1).
•
configure ‘
Input signal value
’ (
InP.S
) to ‘
on
’ to see the input signal
value and the measurement units (for example: ‘
Inp mV 8.52
’). This
value is expressed in millivolts (mV’) (see section 10.1).
•
configure ‘
Output signal value
’ (
out.S
) to ‘
on
’ to see the output signal
value and the measurement units (for example : ‘
Out mA 12.40
’).
•
configure ‘
Label
’ (
LAbL
) to ‘
on
’ to see the value configured at the
‘
label
’ and ‘
label2
’ parameters (see section 13.9).
•
configure ‘
Process value
’ (
Proc
) to ‘
on
’ to see the process value as
scaled at the process parameters (see section 13.4) (for example:
‘
Proc 150.0
’).
•
configure ‘
Percentage
’ (
Prct
) to ‘
on
’ to see the percentage of
input signal, where ‘0’ is the value assigned to the ‘
input signal low
’
parameter, and ‘100’ is the value assigned to the ‘
input signal high
’
parameter (see section 13.4) (for example: ‘
Prct 23.5
’).
•
configure ‘
Excitation voltage
’ (
EX.V
) to ‘
on
’ to see the value of the
excitation voltage received by the load cell. This value is read from the
‘
sense
’ terminals (see section 7.2) (for example : ‘
ExV 4.97
’).
•
configure ‘
Excitation current
’ (
EX.MA
) to ‘
on
’ the value of the current
provided through the excitation voltage terminals (for example: ‘
ExMA
14.3
’).
When more than one parameter is set to ‘
on
’, values will be displayed
sequentially, in the same order as they are listed in the menu, with
a middle dash ‘
-
’ between them. When all information has been
displayed, the instrument returns to ‘
normal mode
’ of operation.
13.8 Key ‘LE’ (‘
messages
’ function)
Key ‘LE’
(messages function)
Input signal value
(in mV’)
Output signal value
Process value
Label
Percentage
Excitation voltage
Excitation current
Tare value (in mV’)
mV at terminals
Summary of Contents for I4 Series
Page 24: ......