background image

F

E

MA

 · MANUFACTURING FOR INDUSTRIAL AUTOMATION 

4

The instrument can be configured to read signals from load cells which 

are externally powered up to 10 Vdc, and not use the power provided by 

the instrument.

Configure the instrument to read in ‘

load cell

’ mode and set the excitation 

voltage parameter to ‘

off

’. Connect the ‘

sense

’ wires to the excitation 

voltage terminals of the load cell. With the ‘

sense

’ wires, the instrument 

will compensate for variations of the power supply.

With this configuration, values indicated in mV’ units (see section 10.1), 

are scaled to a theoretical power value of 5 Vdc, therefor values may not 

be directly interpretable.

The  instrument  accepts  up  to  4  standard  350 Ohms  load  cells.  The 

instrument provides 5 Vdc excitation voltage. For load cells with different 

impedance, calculate the current consumption for each cell, and the 

total must not exceed the maximum current the instrument can provide 

(see section 11).
In case of problems with the signal provided by the load cell, the 

instrument provides information for troubleshooting purposes. Configure 

the ‘

messages

’ function (see section 13.8) to access the actual values 

for the input signal (expressed in mV), the excitation voltage measured 

at the ‘

sense

’ terminals (expressed in Vdc) and the current provided to 

the cell (expressed in mA). The operator can use these values to identify 

the cause of the problem. See section for more information on how to 

access these values in real time.

The instrument is designed to measure load cell signals. The instrument 

provides 5 Vdc excitation voltage to power the load cell, and reads the 

millivolt signal generated by the load cell. The instrument also reads the 

actual excitation voltage connected to the load cell, and compensates 

the read signal for changes at the excitation voltage. 
The actual value of the excitation voltage is detected by using the 

sense

’ wires. Connect the ‘

sense

+

’ and ‘

sense

-

’ (terminals 5 and 2) to 

the load cell, to provide the instrument with an accurate value of the 

excitation voltage received by the cell. Deviations and errors from the 

standard excitation value (5 Vdc) are automatically compensated by the 

instrument, increasing the accuracy and reliability of the measure.
If you can not connect the ‘

sense

’ wires to the load cell, place a 

shortcircuit between terminals ‘

sense

+

’ and ‘

Vexc

+

’ (terminals 5 and 4), 

and between terminals ‘

sense

-

’ and ‘

Vexc

-

’ (terminals 2 and 1).

For applications with multiple load cells (2, 3 or 4 cells) connect the 

sense

 ’ wires to the ‘

electrical middle point

’ of the power wires of all the 

cells (see section 7.8).

The ‘

sense

’ terminals must be always connected. If you do not 

use the ‘

sense

’ wires, shortcircuit with ‘

Vexc

’ terminals

Table 2  

T

ypical load cell connection

Vexc +

signal -

sense -

Vexc -

sense +

1 2 3

4 5 6

I4L terminals

The instrument can be configured to measure millivolts in differential 

mode. Activating any millivolt measurement mode, disables de 

excitation voltage and disables the ‘

sense

’ compensation for changes 

at the excitation voltage. The instrument works as a pure differential 

millivolt signal converter.

7. Practical load cell information

Table 3  

Millivolt mode connection

mV -

mV +

1 2 3

4 5 6

I4L terminals

7.1 Number and type of cells accepted

7.2 Load cell and ‘

sense

’ wires

7.3 Millivolt mode

7.4 Load cell with external power

Measuring with load cells requires an electrically clean installation. 

When connecting the ground to the cell system, assure that the load 

cell connection to ground is performed in such a way that the current to 

ground does not flow through the cell.

7.5 Connecting the cell to the ground

Table 4  

Load cell connection with external power 

signal -

sense -

sense +

1 2 3

4 5 6

Vexc -

Vexc +

I4L terminals

Summary of Contents for I4 Series

Page 1: ...www fema es SERIES I4 Model I4L USER S MANUAL ISOLATED SIGNAL CONVERTER Section INDUSTRIAL 5583r00 INDUSTRIAL Weight Kg Load cell mV...

Page 2: ...ocs 5587_I4L_installation_en pdf CE declaration www fema es docs 5642_CE Declaration_I4_en pdf Warranty www fema es docs 4153_Warranty1_en pdf Web www fema es Series_I4 3 Additional information When t...

Page 3: ...Read the practical load cell information see section 7 4 Connect the input the output and the power terminals see section 9 error messages may appear in the process of connection see section 16 for ex...

Page 4: ...ead signal for changes at the excitation voltage The actual value of the excitation voltage is detected by using the sense wires Connect the sense and sense terminals 5 and 2 to the load cell to provi...

Page 5: ...g 3 load cells is the optimal way to distribute the weight on a plane although it is common to work with 4 load cells in applications with tanks hoppers and similar When working with multiple load cel...

Page 6: ...trument reads uSEr Code identifies the end of the list it follows code 199 and the list continues with code 010 Select to exit the list without applying changes To calculate the optimal input signal r...

Page 7: ...dc 7 8 9 common Vdc fuse signal sense Vexc POWER ABC 18 to 265 Vac dc isolated OUTPUT SIGNAL 789 sense common 0Vdc or passive mA current out signal 4 20mA mA current in signal 0 10Vdc or active mA cur...

Page 8: ...ion voltage terminals OUTPUT SIGNAL The output signal is configurable to 4 20mA active and passive and 0 10Vdc Table 13 Input signal ranges for load cell signals Input range Code for 4 20mA output Cod...

Page 9: ...ifications OUTPUT SIGNAL The output signal is configurable to 4 20mA active and passive and 0 10Vdc Table 14 Connection examples for millivolt signals 1 2 3 not connected mV not connected 4 5 6 not co...

Page 10: ...he input with no filter 115 mSec typ 0 to 99 with 50Hzfilter or 60Hzfilter 150mSec typ 0 to 99 with 50 and 60Hz filter 300mSec typ 0 to 99 OUTPUT SIGNAL RANGES active current output 4 20mA active max...

Page 11: ...SERIES I4 Model I4L Section INDUSTRIAL ISOLATED SIGNAL CONVERTERS www fema es 11 T h i s p a g e b l a n k...

Page 12: ...t returns to normal mode of operation HOW TO OPERATE INSIDE THE CONFIGURATION MENU Inside the configuration menu use the front keypad to move through menu entries parameters and select configuration v...

Page 13: ...ee section 13 8 The messages function ends when all the information has been displayed or front keys UP 5 or SQ are pressed The display returns to normal mode of operation 12 5 How to activate the Mes...

Page 14: ...cted a configuration code see section 13 1 the input signal has been already selected and there is no need to manually select the Mode ModE or Signal range rAnG parameters again At the Mode ModE param...

Page 15: ...ressed in x xx mA or in x xx Vdc at the Output high signal ou hI parameter configure the high output signal value This value is expressed in x xx mA or in x xx Vdc These four parameters define the rel...

Page 16: ...citation current EX MA to read the value of the current provided through the excitation voltage terminals for example ExMA 14 3 Display information Input signal value in mV Output signal value Label S...

Page 17: ...rrected millivolts mV see section 10 1 configure Input signal value InP S to on to see the input signal value and the measurement units for example Inp mV 8 52 This value is expressed in millivolts mV...

Page 18: ...and to underrange output signal to minimum value in case of input signal underrange at the On SQ on Sq parameter configure the behavior of the output signal when the operator is inside configuration...

Page 19: ...set a dead band value of 0 08mVdc All signals below 0 08mVdc will be treated as 0mVdc and the output will be 4mA the Version VEr parameter informs about the firmware version running in the instrument...

Page 20: ...oltage Excitation voltage 0 70mV 0 80mV 10mV 20mV 30mV 40mV 50mV 60mV 70mV 80mV 5mV Output range ok message flashes while the field correction function is being applied and when finished the instrumen...

Page 21: ...abel2 Alphanumerical 0 0 to 100 0 SOS mode of output On error Output to high Output to low Output standard Power filter No filter Filter for 50Hz Filter for 60Hz Filter for 50Hz and 60Hz On SQ Output...

Page 22: ...ell cELL Signal range rAnG 10 00mV Excitation voltage V EXc on Output range out 4 20mA Advanced scaling Ad Sc Tare tArE 0 00 mV Input signal low In Lo 0 00 mV Input signal high In hI 10 00 mV Output s...

Page 23: ...r motor controllers electrovalves together with signal and or control cables The use of shielded cables is recommended to prevent the coupling of environmental electromagnetic noise connected to earth...

Page 24: ......

Reviews: