3.2 Testing the RFID-Tunnel
After tuning the RFID-Tunnel, you can check for proper function using the Reader, the ISOStart
service software and a Smart Tag. Here the Noise Level and performance of the gates are tested.
3.2.1 Checking the Noise Level
Step
Action
Note
1
Select with “Commands”
of multiplexer
Gate 1 (Out 1+8) or
Gate 2 (Out 2+7) or
Gate 3 (Out 3+6 )
Cascade Level 1
2
Confirm with “Send”
3
Activate “Test and
Measurement”
4
Select „Noise Level“ and
start by clicking on „Start“
5
Normal Noise Level values:
Average:
< 50mV
Difference (Max-Min):
< 30mV
If the values are not met, check the following:
•
Are all cables pulled tight and do they make good contact?
•
Were the ring cores installed in the antenna cable?
•
Were the cables routed as specified?
•
Are other RFID systems installed nearby?