8.2. KSW Custom Commands
These commands supports the functions of the KSW TempSens and VarioSense transponder
8.2.1. [0xA0] Set Passive
This command deactivates the RC-oscillator of the transponder. The temperature detection will be
switched off.
REQUEST-DATA
5
6
(7-14)
0xA0
MODE
UID
RESPONSE-DATA (STATUS = 0x95)
5
ISO-ERROR
MODE:
Bit:
7
6
5
4
3
2
1
0
Function
-
-
-
-
-
ADR
ADR:
ADR
b000
non-addressed
b001
addressed
b010
selected
UID:
Read-only serial no. of the transponder. The UID is required only in the addressed mode.
ISO-ERROR: