ST100 Series Configuration Software

10 

Fluid Components International LLC 

Figure 18 – Example 4-20mA User Tab (Configuration)

The table below summarizes the tabs within the 

Configuration

 branch. 

Table 3 – Configuration Tabs

Tab Name

Tab Description

Password Level

Output

Analog Output board

: Sets 4-20 mA channel configuration

1

 and Frequency 

and Pulse output configuration. 

Digital Output board

: Selects Modbus or FF/Profibus)

2

User

4-20mA User

Manual mA Output loop check; configure/enable NAMUR fault.

User

Modbus

Sets Modbus communication parameters.

User

Extended Op. Mode

Expands flow measurement capabilities by providing 3 additional modes 

of operation. Refer to 

Extended Operation Modes

 in main manual 

06EN003400

 for detailed information.

User

Group Switch Setup

Sets up automatic calibration group switching depending on an external 

4-20 mA output driving the ST100 auxiliary input port. Refer to 

External 

Control Group Switching (EGS)

 in main manual 

06EN003400

 for detailed 

information.

User

AST Power Mode

For ST100 units with AST only

:  Sets heater mode (Constant Delta T 

[AST] or Constant Power) and max. heater current for AST (90 mA or 105 

mA). The max. current value forms the threshold at which the instrument 

switches over to Constant Power mode.

User

Note 1. To set 

HART

 operation, select 

HART Flow

 from 4-20 mA #1 drop-down list (in 

Analog Output Selection

 field).

Note 2. Digital busses (includes HART, Modbus, and FF/Profibus) are mutually exclusive, meaning only one can be active at 

a time. Attempting to enable HART when Modbus or FF/Profibus is in effect causes the Digital Bus Deactivation Warning 

dialog to display: Click 

OK

 to make the change and force the Digital Output Selection to 

None

 or click 

Cancel

 to leave the 

setting unchanged. Attempting to enable Modbus or FF/Profibus when HART is in effect causes the  HART Deactivation 

Warning dialog to display: Click 

OK

 to make the change and force the 4-20 mA #1 Selection to 

Flow

 or click 

Cancel

 to leave 

the setting unchanged. 
[User password 2772]

To verify the current configuration of any setup parameter, click 

Get from Device

 on any of the Setup menus. After changing any of the setup 

parameters, click 

Send to Device

. Click 

Get from Device

 again to verify the parameter(s) change. Observe that the changed parameters are now 

displayed. The remaining 

Configuration

 tab screens are shown below.

Summary of Contents for ST100 series

Page 1: ...ST100 Series Thermal Mass Flow Meter Configuration Software Manual Fluid Components International LLC FCI All rights reserved...

Page 2: ...al LLC FCI Disclosure of this data to you is expressly conditioned upon your assent that its use is limited to use within your company only and does not include manufacture or processing uses Any othe...

Page 3: ...Ethernet Tab Advanced Setup 8 Figure 14 Example Data and Time Tab Advanced Setup 8 Figure 15 Example Download Calibration Tab Advanced Setup 8 Figure 16 Example Reboot Device Tab Advanced Setup 9 Fig...

Page 4: ...Series Configuration Software iv Fluid Components International LLC List of Tables Table 1 Basic Setup Tabs 4 Table 2 Advanced Setup Tabs 7 Table 3 Configuration Tabs 10 Table 4 Diagnostics Tabs 13 T...

Page 5: ...application Use Ethernet for remote applications in which the host PC communicates with the instrument over an Ethernet network Refer to Advanced Setup Tab Screens on page 7 for info on setting Ethern...

Page 6: ...HMI front panel display includes the following Flow as percent of range scale Flow with engineering units Total Flow if Mass or Volumetric units used Temperature Pressure for STP Series only Calibrati...

Page 7: ...Get from Device and or Send to Device buttons at the bottom portion of the window These buttons are shown if the window tab includes parameter data that can be retrieved from the instrument for displ...

Page 8: ...ed no pass word required User Units Select flow and temperature units User Pipe Size Select pipe type and dimensions User Alarms Select and set alarm requirements User SD Card Logging Select logging r...

Page 9: ...ST100 Series Configuration Software Fluid Components International LLC 5 Figure 6 Example Units Tab Basic Setup Figure 7 Example Pipe Size Tab Basic Setup Figure 8 Example Alarms Tab Basic Setup...

Page 10: ...ries Configuration Software 6 Fluid Components International LLC Figure 9 Example SD Card Logging Tab Basic Setup Figure 10 Example Totalizer Tab Basic Setup Figure 11 Example Pressure Offset Tab Basi...

Page 11: ...lock date and time Lets users copy the system host PC time and transmit it to the instrument s battery backed real time clock User Download Calibration Lets users download a full calibration to their...

Page 12: ...onfiguration Software 8 Fluid Components International LLC Figure 13 Example Ethernet Tab Advanced Setup Figure 14 Example Data and Time Tab Advanced Setup Figure 15 Example Download Calibration Tab A...

Page 13: ...branch on the menu tree to access configuration setup items The Output tab is the first of several tabs across the top of the screen Each tab provides a particular menu within the Configuration branch...

Page 14: ...T AST or Constant Power and max heater current for AST 90 mA or 105 mA The max current value forms the threshold at which the instrument switches over to Constant Power mode User Note 1 To set HART op...

Page 15: ...Configuration Software Fluid Components International LLC 11 Figure 19 Example Modbus Tab Configuration Figure 20 Example Extended Op Mode Tab Configuration Figure 21 Example Group Switch Setup Tab C...

Page 16: ...ST100 Series Configuration Software 12 Fluid Components International LLC Figure 22 Example AST Power Mode Tab Configuration...

Page 17: ...history Click Get Fault Logs from Device to list the faults in the scrollable text box Click Clear Fault Log to clear the log User idR Scheduled Tests1 For internal Delta R idR resistance check Selec...

Page 18: ...ational LLC Figure 24 Example Fault Log Tab and Example Fault Log List Diagnostics Figure 25 Example idR Scheduled Tests Tab and Example idR On Demand Test Results Display Diagnostics Figure 26 Exampl...

Page 19: ...tory use only 4 20 mA output DAC count scaling and manual output con trol plus 4 20 mA input raw A D counts and gain offset adjustment Factory Options Factory use only Option inventory HMI display FEs...

Page 20: ...ST100 Series Configuration Software 16 Fluid Components International LLC Figure 29 Example 4 20mA Factory Tab Factory Figure 30 Example Options Tab Factory Figure 31 Example HART Tab Factory...

Page 21: ...ST100 Series Configuration Software Fluid Components International LLC 17 Figure 33 Example Reset idRs Tab Factory Figure 32 Example Memory Tab Factory...

Page 22: ...eal time pressure value applies to STP models only This screen can be helpful when diagnosing system faults FE1 FE2 Process Data Depending on the system configuration options the application menu tree...

Page 23: ...r a particular calibration group numbered 1 5 Selecting a parameter report for a particular calibration group displays that group s info data Similar to other setup menus there is a Send Changes to De...

Page 24: ...able by facsimile if all possibilities have been exhausted with the authorized factory representative Our fax number is 1 760 736 6250 it is available 7 days a week 24 hours a day By Email FCI Custome...

Page 25: ...ST100 Series Configuration Software Fluid Components International LLC 21 NOTES...

Page 26: ...hibited without the prior written consent of FCI Visit FCI on the Worldwide Web www fluidcomponents com FCI World Headquarters 1755 La Costa Meadows Drive San Marcos California 92078 USA Phone 760 744...

Reviews: