F-30-31-35.XX_User Manual_Rev.6 _ 07.11.2012 CORREZIONI 

 

16 / 28

 

 

1.

 

DÉTACHER LES PARTIES JETABLES ET LES ÉLIMINER. 

2.

 

S'ASSURER QUE L'ASPIRATEUR EST EN PARFAIT ÉTAT. 

3.

 

PROCÉDER À SON NETTOYAGE COMME INDIQUÉ CI-APRÈS. 

4.

 

LE RECHARGER. 

IL EST OPPORTUN DE STÉRILISER LES TUYAUX D'ASPIRATION APRÈS CHAQUE UTILISATION. 

LE FILTRE EST JETABLE; IL FAUT DONC LE REMPLACER CHAQUE FOIS QU'ON S'EST SERVI DE L'APPAREIL ET S'ASSURER, AVANT DE 
L'UTILISER À NOUVEAU, QUE CE REMPLACEMENT A ÉTÉ FAIT 

NETTOYAGE DU BOCAL 

LE  BOCAL PEUT ÊTRE  MIS  EN AUTOCLAVE À  UNE  TEMPÉRATURE  N'EXCÉDANT PAS  121  ºC    POUR  UN  MAXIMUM  DE  3  MINUTES  1  ATM, 
VAPEUR SUR LE NAVIRE PEUT "ETRE ABIME, AU-DELÀ DESQUELLES IL RISQUE D'ÊTRE ENDOMMAGÉ- 
IL  PEUT  AUSSI  ÊTRE  DÉSINFECTÉ  AVEC  DE  L'ALCOOL  DÉNATURÉ  OU  UNE  SOLUTION  À  BASE  D'HYPOCHLORITE  EN  VENTE  DANS 
TOUTES LES PHARMACIES.  LIMITE MAXIMUM DE RÉUTILISATION: 30 CYCLES. 

NETTOYAGE DE L'ASPIRATEUR 

UNE FOIS L'APPAREIL DÉBRANCHÉ, LE NETTOYER AVEC UN CHIFFON DOUX IMBIBÉ DE DÉTERGENT NON ABRASIF. 

CONTRÔLE PÉRIODIQUE D'ENTRETIEN 

L'APPAREIL EST CONÇU DE FAÇON QUE LES OPÉRATIONS D'ENTRETIEN SOIENT LIMITÉES A UN MINIMUM. IL EST CEPENDANT INDISPENSABLE DE LE 
CONTRÔLER TOUS  LES  JOURS  AVANT  DE  S'EN  SERVIR.  AVANT  DE  VÉRIFIER  QUOI  QUE  CE  SOIT, EN  CAS  D'ANOMALIE  OU  DE  FONCTIONNEMENT 
DÉFECTUEUX, PRENDRE CONTACT AVEC LE SERVICE TECHNIQUE FAZZINI S.R.L. OU AVEC UN TECHNCIEN AUTORISÉ PAR CETTE SOCIÉTÉ.

  

OPERATION QUI DOIT ETRE EXECUTEE EXCLUSIVEMENT PAR UN PERSONNEL QUALIFIE. 
FAZZINI.  S.R.L.  NE  GARANTIT  PAS  LES  APPAREILS  DONT  SON  SERVICE  TECHNIQUE  AURA,  LORS  D'UN  CONTRÔLE,  CONSTATÉ  QU'ILS  ONT  ÉTÉ 
INDÛMENT  MANIPULÉS.  L'OPÉRATEUR  DOIT  EXCLUSIVEMENT  SE  LIMITER  À  REMPLACER,  LE  CAS  ÉCHÉANT,  LES  ACCESSOIRES  FOURNIS  EN 
DOTATION, DANS LA MESURE DE L'USAGE QU'IL FAIT DE L'APPAREIL. 
IL EST OPPORTUN DE STÉRILISER LES TUYAUX D'ASPIRATION APRÈS CHAQUE UTILISATION. 
AU  CAS  OÙ  UNE  RÉPARATION  SERAIT  NÉCESSAIRE,  SE  METTRE  EN  CONTACT  AVEC  FAZZINI  S.R.L.  LE  PAIEMENT  DES  FRAIS  D'EXPÉDITION  DE 
L'APPAREIL  ENDOMMAGÉ  DOIT  ÊTRE  ÉTABLI  EN  ACCORD  AVEC  LA  SOCIÉTÉ  AVANT  QUE  CELUI-CI  NE  LUI  SOIT  RENVOYÉ.  DANS  LE  CAS  DE 
PÉNÉTRATION DE LIQUIDES À L'INTÉRIEUR DE L'APPAREIL, SE METTRE EN CONTACT AVEC LE SERVICE TECHNIQUE DE FAZZINI S:R.L. POUR LE FAIRE 
RÉPARER. NE PAS INTERVENIR DIRECTEMENT SUR L'APPAREIL. 

CONSERVATION 

L'ASPIRATEUR DOIT ÊTRE RANGÉ DANS UN ENDROIT SEC, À L'ABRI DE L'ACTION DES AGENTS ATMOSPHÉRIQUES. 
TEMPÉRATURE DE CONSERVATION: ENTRE - 10° ET + 55°. HUMIDITÉ RELATIVE: ENTRE 20 ET 80%. 

CHARGE DEL F-31.XX 

1.

 

LORSQUE LE DERNIER LED ROUGE RESTE ALLUME SUR L’ECRAN DE L’APPAREIL, CONTRAIREMENT AU VERT ET AUX DEUX JAUNES QUI 
SONT ETEINTS, INTERROMPRE LE FONCTIONNEMENT DE L’APPAREIL ET LE FAIRE CHARGER. 

2.

 

POUR FAIRE CHARGER L’APPAREIL PROCEDER COMME SUIT : INTRODUIRE LA FICHE A L’ARRIERE SUR 230 V. LE LED INTERMITTENT 
S’ALLUMERA. DEVIER L’INTERRUPTEUR SUR «  CHARGE » ET FAIRE CHARGER PENDANT 12 HEURES AU MOINS 

L'APPAREIL PEUT FONCTIONNER AVEC UNE BATTERIE EXTERNE 12 V,  AVEC UN CABLE PREVU A CET EFFET ET UNE PRISE VOITURE  
COMMENT REPLACER LA BATTERIE: 
ETEINDRE L’ASPIRATEUR ET COUPER LA COURANT 
DEVISSER LE 4 VIS ARRIERES (2 A COTE DU MANCHE ET 2 SOUS) 
LA BATTERIE EST FIXEE EN SERRE-JOINT DEDANS L’ASPIRATEUR 
DEVISSER LES 2 VIS DU SERRE-JOINT 
SORTIR LA BATTERIE ET RACCORDER LA NOUVELLE SUIVANT LES POLARITES APPROPRIEES 
POLARITE POSITIVE (+) CABLE BLEU OU ROUGE 
POLARITE NEGATIVE (-) CABLE MARRON  
VISSER LES 2 VIS DU SERRE-JOINT 
FERMER L’ASPIRATEUR EN VISSANT LES 4 VIS 

ÉLIMINATION 

POUR L'ÉLIMINATION DE L'APPAREIL, DE SES ACCESSOIRES ET DES PARTIES JETABLES CONTAMINÉES, SE CONFORMER AUX NORMES EN VIGUEUR 
EN MATIÈRE D'ÉLIMINATION DES DÉCHETS. FAZZINI S.R.L. SE RÉSERVE TOUS LES DROITS DE MODIFIER LA PRÉSENTE NOTICE SANS PRÉAVIS, PAR 
SUITE DE MODIFICATIONS APPORTÉES À LA CONSTRUCTION DE L'APPAREIL QUI A ÉTÉ FABRIQUÉ CONFORMÉMENT AUX NORMES SUIVANTES: 
- CEI EN 60601.1 (CEI 62.5) / IEC 601.1 / UNI EN ISO 10079/1:00 - MARQUE CE DE CONFORMITÉ À LA DIRECTIVE   93/42 CEE 
LA SOCIÉTÉ FAZZINI S.R.L. S'IMPOSE LE MONITORAGE DES APPAREILS MIS SUR LE MARCHÉ, AUSSI VOUS PRIE-T-ELLE DE BIEN VOULOIR LUI 
SIGNALER TOUS LES PROBLÈMES QUI POURRAIENT SE PRÉSENTER OU LES SUGGESTIONS QUE VOUS INSPIRENT NOS PRODUITS, EN ENVOYANT VOS 
COMMUNICATIONS AUX ADRESSES SUIVANTES: 
 
FAZZINI SRL - S.S. PADANA SUP. 317- 20090 VIMODRONE (MI) – ITALIA - TÉLÉCOPIE: 0039/0227409242 - E-MAIL [email protected] 
 
VEUILLEZ PRÉCISER QUE VOUS AUTORISEZ LE TRAITEMENT DES DONNÉES AUX TERMES DE LA LOI 196/03. 

Summary of Contents for F-31

Page 1: ...ure attentive de cette notice explicative Vous y trouverez toutes les informations nécessaires pour une utilisation correcte de l appareil Vous êtes priés de suivre attentivement les indications qu elle contient et de la conserver inaltérée dans un endroit facilement accessible à l opérateur Le fabricant se réserve le droit de modifier le contenu de cette notice à n importe à quel moment et sans p...

Page 2: ...F 30 XX F 31 XX F 35 XX Istruzioni per l uso Instructions for use Mode d emploi Manual de instrucciones Instruções de uso F 30 31 35 XX_User Manual_Rev 6 _ 07 11 2012 CORREZIONI 2 28 ...

Page 3: ...olon jar Bocal en makrolon Frasco en makrolon 1000 ml 05 0001 Filtro monouso Filter disposable Filtre usage unique Filtro desechable Filtro monouso 05 0016 Sonda d aspirazione sterile monouso Suction curette disposable Sonde d aspiration usage unique Tubo de aspiracion desechable CH16 Reorder Code 5 0010 Kit tubi monouso Set of disposable tubes Jeu de tuyaux usage unique Jogo de tubos no reutiliza...

Page 4: ... per l aspirazione di fluidi organici nei trattamenti ospedalieri e ambulatoriali E indispensabile verificare le caratteristiche elettriche e di protezione IP riportate nei dati di targa prima di utilizzare l aspiratore Attenzione Non aspirare gas quali ossigeno o sue miscele L utilizzo improprio di questo apparecchio può essere pericoloso per il paziente INSTRUCTIONS FOR USE Duty time 30 min ON 3...

Page 5: ...as comme l oxygène et ses mélanges L utilisation incorrecte de cet appareil peut être dangereuse pour le patient INSTRUCCIONES MÁS DESTACADAS Ciclo de funcionamiento 30 min ON 30 min OFF REGULADOR DE VACÍO El regulador de vacío que se puede identificar por medio de la ruedilla verde permite regular el circuito del aire Si hace girar la ruedilla hacia la derecha la presión negativa aumentará hasta ...

Page 6: ...OSTO A ALTRI APPARECCHI NEL CASO E NECESSARIO OSSERVARE L APPARECCHIO PER CONTROLLARE IL FUNZIONAMENTO NORMALE NELLA CONFIGURAZIONE IN CUI E USATO BATTERIA SOLO PER MODELLO F 31 XX ATTENZIONE MANTENERE SOTTO CARICA L APPARECCHIO QUANDO NON È UTILIZZATO LA RICARICA PUÒ ESSERE EFFETTUATA CON L APPOSITO CAVO ESTERNO O INSERENDO L APPARECCHIO SULLA APPOSITA STAFFA DI RICARICA PRIMA DELL UTILIZZO INIZI...

Page 7: ...Istruzioni per l uso Instructions for use Mode d emploi Manual de instrucciones Instruções de uso F 30 31 35 XX_User Manual_Rev 6 _ 07 11 2012 CORREZIONI 7 28 CARATTERISTICHE TECNICHE BATTERIA SOLO PER F 31 XX 45 MINUTI CIRCA ...

Page 8: ...ETTONE SENZA GALLEGGIANTE 5 COLLEGARE L ALTRA ESTREMITÀ DEL TUBO LUNGO 4 AL RACCORDO DELLA SONDA 6 IN DOTAZIONE 6 COLLEGARE LA SPINA ALLA PRESA DI RETE ELETTRICA SOLO DOPO AVER CONTROLLATO CHE LA TENSIONE IN RETE SIA EQUIVALENTE ALLA TENSIONE DI FUNZIONAMENTO RIPORTATA SULLA TARGHETTA POSTA SULLA PARTE POSTERIORE DELL APPARECCHIO 7 PREMERE L INTERUTTORE ON OFF 7 8 L INTENSITA DI ASPIRAZIONE PUO ES...

Page 9: ...ILIZZO NON UTILIZZARE L ASPIRATORE PRIVO DEL FILTRO PULIZIA VASO TUBI ATTENZIONE IL VASO RACCOLTA LIQUIDI E I TUBI POSSONO ESSERE MESSI IN AUTOCLAVE AD UNA TEMPERATURA NON SUPERIORE A 121 ºC PER UN PERIODO MASSIMO DI 3 MINUTI DI 3 MINUTI 1 ATM VAPORE SATURO OLTRE IL VASO POTRA RISULTARE DANNEGGIATO OLTRE IL VASO POTRA RISULTARE DANNEGGIATO IN ALTERNATIVA DISINFETTARE UTILIZZANDO ALCOOL DENATURATO ...

Page 10: ...OSTE AI LATI DEL MANIGLIONE E LE ALTRE 2 POSTE SOTTO APRENDO L APPARECCHIO SI TROVERÀ LA BATTERIA FISSATA TRAMITE STAFFA SVITARE LE 2 VITI CHE FISSANO LA STAFFA SCOLLEGARE LA BATTERIA DA SOSTITUIRE E COLLEGARE QUELLA NUOVA SEGUENDO LE POLARITÀ CORRETTA POSITIVO FILO BLU O ROSSO NEGATIVO FILO MARRONE RIFISSARE CON LE DUE VITI LA STAFFA CHE TIENE LA BATTERIA RICHIUDERE L ASPIRATORE RIAVVITANDO LE 4 ...

Page 11: ...HE FIRST TIME IT IS NECESSARY TO KEEP BATTERIES UNDER CHARGE FOR 24 HOURS NON STOP A BATTERY CONSTANTLY UNDER CHARGE LASTS ABOUT 2 OR 3 YEARS AFTER THIS PERIOD OF TIME IT IS RECOMMENDED TO REPLACE IT IF THE APPLIANCE CANNOT BE PUT UNDER CONTINUOUS RECHARGE RECHARGE IT FOR 24 HOURS EVERY 15 DAYS WHEN FINISHED USING IT THE DEVICE MUST ALWAYS BE RECHARGED ATTENTION THE BATTERY MAY BE DAMAGED IF NOT R...

Page 12: ... 11 IMMEDIATELY SWITCH OFF THE APPLIANCE BY SETTING THE ON OFF 7 SWITCH ON THE OFF POSITION INDICATORS LED ONLY FOR MODEL F 31 XX THE ASPIRATOR F 31 XX IS EQUIPPED WITH LIGHT INDICATORS LED SIGNALING THE BATTERY USE BECAUSE USING THE APPLIANCE ON A LOW BATTERY MAY DAMAGE THE APPLIANCE ITSELF WHILE CONTINUOUSLY RECHARGING IT DOES NOT CAUSE ANY DAMAGE IT IS RECOMMENDED TO ALWAYS CHECK THE BATTERY S ...

Page 13: ...COSTS MUST ALWAYS BE AGREED WITH FAZZINI S R L BEFORE SHIPPING FAILED PIECES SHOULD ANY FLUID LEAK IN THE APPLIANCE CALL THE TECHNICAL SERVICE OF FAZZINI S R L FOR MAINTENANCE DO NOT OPEN THE UNIT STORAGE THE ASPIRATOR MUST BE STORED IN A DRY LOCATION PROTECTED FROM ATMOSPHERIC AGENTS STORAGE TEMPERATURE BETWEEN 10 AND 55 RELATIVE HUMIDITY BETWEEN 20 AND 80 TO CHARGE DEVICE F 31 XX 1 WHEN ON THE D...

Page 14: ... CHARGEE DURE ENVIRON 2 OU 3 ANS APRES QUOI IL EST OPPORTUN DE LA REMPLACER SI L APPAREIL NE PEUT ETRE RECHARGE DE FAÇON CONTINUE PREVOIR UNE RECHARGE DE 24 H TOUS LES 15 JOURS L APPAREIL DOIT TOUJOURS ETRE RECHARGE APRES L USAGE LA BATTERIE RISQUE DE S ENDOMMAGER SI CETTE PRÉCAUTION N EST PAS OBSERVÉE UNE BATTERIE COMPLETEMENT DECHARGEE QUI N A PAS ETE IMMEDIATEMENT RECHARGEE APRES L UTILISATION ...

Page 15: ...QUANT L ÉTAT DE LA BATTERIE SI ON L UTILISE AVEC UNE BATTERIE DÉCHARGÉE L ASPIRATEUR PEUT ÊTRE GRAVEMENT ENDOMMAGÉ CE QUI NE RISQUE PAS DE SE PRODUIRE SI ON LE RECHARGE CONTINUELLEMENT NOUS VOUS INVITONS PAR CONSÉQUENT À TOUJOURS BIEN VÉRIFIER L ÉTAT DE LA BATTERIE SIGNAUX LUMINEUX DESCRIPTION DU NIVEAU DE RECHARGE LEDS ALLUMÉS VERT NIVEAU MAX D AUTONOMIE 2 LEDS ALLUMÉS JAUNE 50 AUTONOMIE MOYENNE ...

Page 16: ... CI NE LUI SOIT RENVOYÉ DANS LE CAS DE PÉNÉTRATION DE LIQUIDES À L INTÉRIEUR DE L APPAREIL SE METTRE EN CONTACT AVEC LE SERVICE TECHNIQUE DE FAZZINI S R L POUR LE FAIRE RÉPARER NE PAS INTERVENIR DIRECTEMENT SUR L APPAREIL CONSERVATION L ASPIRATEUR DOIT ÊTRE RANGÉ DANS UN ENDROIT SEC À L ABRI DE L ACTION DES AGENTS ATMOSPHÉRIQUES TEMPÉRATURE DE CONSERVATION ENTRE 10 ET 55 HUMIDITÉ RELATIVE ENTRE 20...

Page 17: ...E SE RESEÑAN EN EL PRESENTE MANUAL PUEDE MENOSCABAR LA SEGURIDAD Y LOS PARÁMETROS TÉCNICOS DEL MISMO BATERÍA SOLO PARA F 31 XX CUIDADO MANTENGA ENCHUFADO EL APARATO MIENTRAS NO LO UTILIZA PARA QUE SE CARGUE EL MISMO SE PUEDE RECARGAR MEDIANTE EL CABLE EXTERIOR AL EFECTO O BIEN INTRODUCIENDO EL APARATO EN EL ESTRIBO DE RECARGA AD HOC ANTES DE USARLO POR PRIMERA VEZ ES MENESTER CARGAR LAS BATERÍAS D...

Page 18: ... FLOTANTE 5 CONECTE EL OTRO EXTREMO DEL TUBO LARGO 4 CON EL RACOR 6 A FIN DE ACOPLAR LA SONDA Y EN SU CASO LA SONDA QUE VIENE CON EL EQUIPAMIENTO DE SERIE 6 CONTROLE QUE LA TENSIÓN DE LA RED CORRESPONDA A LA TENSIÓN DE FUNCIONAMIENTO QUE SE INDICA EN LA PLACA QUE ESTÁ UBICADA EN LA PARTE TRASERA DEL ARTEFACTO ACTO SEGUIDO ENCHÚFELO A LA RED ELÉCTRICA 7 PULSE EL INTERRUPTOR ON OFF 7 8 LA INTENSIDAD...

Page 19: ...PUEDE ESTERILIZAR EN AUTOCLAVE A UNA TEMPERATURA NO SUPERIOR A LOS 121 ºC DURANTE NO MÁS DE 3 MINUTOS 1 ATM VAPOR SOBRE EL BUQUE SE PUEDE DAÑAR SI SE SUPERA DICHO LAPSO EL RECIPIENTE SE PUEDE MENOSCABAR COMO ALTERNATIVA EL RECIPIENTE SE PUEDE DESINFECTAR CON ALCOHOL DESNATURALIZADO O MEDIANTE UNA DISOLUCIÓN A BASE DE HIPOCLORITO DE VENTA EN LAS FARMACIAS LÍMITE MÁXIMO DE REUTILIZACIÓN 30 CICLOS LI...

Page 20: ...E LOS 10 C Y LOS 55 C LA HUMEDAD RELATIVA HA DE OSCILAR ENTRE EL 20 Y EL 80 ELIMINACIÓN DEL APARATO A LA HORA DE ELIMINAR EL APARATO SUS ACCESORIOS O LAS PARTES DESECHABLES DEL MISMO EL USUARIO DEBE ATENERSE A TODO LO ESTABLECIDO EN LAS NORMATIVAS VIGENTES EN MATERIA DE ELIMINACIÓN DE RESIDUOS FAZZINI S R L SE RESERVA TODOS LOS DERECHOS DE MODIFICAR EL PRESENTE MANUAL SIN PREAVISO CADA VEZ QUE SE ...

Page 21: ...CISO CARREGA LO DURANTE 24 HORAS CONSECUTIVAS AS PILHAS CONSTANTEMENTE CARREGADAS DURAM DE 2 A 3 ANOS DEPOIS É PRECISO TROCA LAS SE NÃO TIVER A POSSIBILIDADE DE CARREGAR O APARELHO CONSTANTEMENTE È PRECISO RECARREGA LO 24 HORAS SEGUIDAS A CADA 15 DIAS APÓS O USO RECARREGAR SEMPRE O APARELHO ATENÇÃO SE NÃO FOREM RECARREGAR APÓS O USO AS PILHAS PODEM FICAR DANIFICADAS AS PILHAS COMPLETAMENTE DESCARR...

Page 22: ...DOR QUE SE ENCONTRA NA PARTE ANTERIOR DO APARELHO 8 E PODERÁ SER CONTROLADA NO DISPOSITIVO QUE SE ENCONTRA NA PARTE DA FRENTE DO VACUÓMETRO 9 9 ACONSELHA SE REALIZAR O PRIMEIRO CICLO DE ASPIRAÇÃO COM PELO MENOS 500 ML DE ÁGUA SOMENTE 10 O LÍQUIDO ASPIRADO ENTRA NO FRASCO SE A QUANTIDADE ULTRAPASSAR A CAPACIDADE DO FRASCO A VÁLVULA QUE CONTROLA O NÍVEL DO LÍQUIDO INTERROMPE A ASPIRAÇÃO 11 DESLIGAR ...

Page 23: ...FUNCIONAMENTO CONTACTAR O SERVIÇO TÉCNICO DA FAZZINI S R L OU OS ESPECIALISTAS DA EMPRESA AS PILHAS DEVEM SER TROCADAS SOMENTE POR PESSOAL COMPETENTE A FAZZINI S R L NÃO GARANTE APARELHOS QUE APÓS O CONTROLO DO SERVIÇO TÉCNICO CONSTAM ALTERADOS O UTILIZADOR SÓ DEVERÁ TROCAR QUANDO FOR NECESSÁRIO OS ACESSÓRIOS FORNECIDOS COM O APARELHO DE ACORDO COM A INTENSIDADE DE SUA UTILIZAÇÃO O UTILIZADOR DEVE...

Page 24: ...LTER WATER REPELLENT ANTIBACTERIAL KIT W CONNECTOR B 05 0001 1 LITER AUTOCLAVABLE JAR AT 121 C COMPLETE W COVER C SP 30 00 JAR SUPPORT D 05 00552 DISPOSABLE SET OF SHORT AND LONG TUBE E 05 00063 CONNECTOR BETWEEN SILICON TUBE AND CATHETER CONNECTOR F SP 38 02 12V DC PLUG G SP 38 06 12V POWER CABLE H SP 34 01 240V AC PLUG I 03 2507 POWER CABLE J N 09 000 01 ROCKER SWITCH K SP 16 01 VACUUM GAUGE L S...

Page 25: ...ustomer or the user of the equipment should assure that it is used in such an environment Immunity test IEC 60601 Test level Compliance level Electromagnetic environment guidance Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 3...

Page 26: ...site survey should be considered If the measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above the equipment should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the equipment b Over the frequency range 150 kHz to 80 MHz field strength...

Page 27: ... EQUIPMENT CLASS 2 ELECTRIC APPAREIL DE CLASSE 2 APARATO DE CLASSE 2 APARELHO CLASSE 2 FUSIBILE FUSE FUSIBLE FUSIBLE FUSÍVEL TENSIONE PERICOLOSA DANGEROUS VOLTAGE TENSION DANGEREUSE VOLTAJE PELIGROSO TENSÃO PERIGOSA IP XX GRADO DI PROTEZIONE CONTRO LA PENETRAZIONE DEI LIQUIDI E SOLIDI DEGREE OF PROTECTION AGAINST LIQUID AND SOLID PENETRATION DEGRÉ DE PROTECTION CONTRE LA PÉNÉTRATION DE LIQUIDES GR...

Page 28: ... il n est plus utilisé n est pas un déchét urbain mais il doit être écoulé avec la collection séparée autorisée Decreto WEEE 2002 95 EC 200296 CE 2003 108 CE el viejo aparato no está residuos municipales deben ser eliminados por mí desechar de forma distinta autorizado Decreto REEE 2002 95 CE 200296 CE 2003 108 CE o aparelho velho não é lixo municipal devem ser eliminados por mim descartados separ...

Reviews: