F H 8 4 0 0
42
42
42
42
TEC 07.07.04/ 412521
Revision index F
08/07/2013
Transmission unit
for the temperature
"115 Temperature" menu
Possible units: °C; °F; K
Setting the analog outputs
Allocation
of the output n° 1
"1200 Set 04-20 Var" menu
The output n° 1 can be allocated to one of these variables:
−
Velocity of sound
−
Velocity of flow
−
Flow
−
Totalizer
−
Pressure
−
Temperature
−
Gain
The gain corresponds to the average value, for the
paths in operation, of the attenuation of ultrasonic
waves through the fluid.
The acceptable values go from -2 dB (not very
attenuated signal; risk of saturation above) to -45 dB
(strongly attenuated signal; risk of interference below)
Calibration of 4 and 20 mA limits
of the analog output n° 1
"1201 Set 04 mA Value" and "1202 Set 20 mA Value" menus
Summary of Contents for FH8400
Page 1: ...FH 8300 1 ...
Page 2: ......
Page 6: ...FH 8400 4 4 4 4 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 16: ...FH 8400 14 14 14 14 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 26: ...FH 8400 24 24 24 24 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 35: ...Chapter 2 Commissioning procedure 33 33 33 33 Connecting the FH8400 analog 4 20 mA output ...
Page 38: ...FH 8400 36 36 36 36 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 50: ...F H 8400 48 48 48 48 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 64: ...F H 8400 62 62 62 62 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 92: ...F H 8400 90 90 90 90 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 94: ...F H 8400 92 92 92 92 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 98: ...F H 8400 96 96 96 96 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 104: ...F H 8400 102 102 102 102 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 110: ...F H 8400 108 108 108 108 TEC 07 07 04 412521 Revision index F 08 07 2013 ...
Page 111: ...109 109 109 109 Personal notes Personal notes Personal notes Personal notes ...