Faro EVA User Manual Download Page 9

EVA           

 

USER MANUAL - EN 

Edition 1.0 April 2020    

 

 

 

 

 

 

                                      Pag. 9 di 14 

              

 

4.9

 

SYNCRO MODE WITH FARO ROOM LIGHT 

When present, the device can be connected by Radio Frequency (RF) connection to the Faro Room Light to 

create a synchronized lighting system. 
The procedure to create this connection is called “pairing”. 

If more than one Room Lights are installed in the cabinet, take care that the other Room Lights are turned 

off or are turned on for more than 60 seconds. 
To activate the 

“Pairing”

, proceed as follows: 

1. Switch on the Faro Room Light. The Room Light will start Searching an input from the dental light for 60 

seconds. 
2. Within the 60 seconds, press the “Pairing” button on the dental lamp between 4 and 6 seconds. But not 

for more than 6 seconds ( otherwise the procedure will be cancelled). 

The Room Light activates the blue LED on the aluminium body.  

If the blue LED does not switches on, it is possible another attempt (within the 60 seconds by 

the firsts). If 60 seconds past from the switching on of the Room Light, it is necessary to repeat 
the procedure from point 1. 

3. After the blue LED switches on the room light, there are 60 seconds to confirm the 

“Pairing” 

by pressing 

the programming button placed on the remote control of the Room Light. At this point, the blue LED of the 
Room Light blinks at double frequency and then switches off. If the button on the remote controls is not 

pressed within 60 seconds, the blue LED switch off and the procedure must be repeated from point 1. 

After the 

“Pairing”, 

synchronization between the 2 lamps (dental light and room light) is enabled. 

To 

DE-ACTIVATE THE SYNCHRONIZATION FUNCTION

, proceed as follows: 

Press the Synchro button from 2 to 4 second, then release it. 
On release, a sound signal (Beep) will be heard and the blue LED on the Room Light lamp will go out to 

indicate that synchronization has been disabled.

 

 

 

When the Faro Room Light is synchronized with the Dental Light the blue LED on the room light is lighted in steady mode. If the LED is off, the 

synchronization is not active. 
The remote control is always enabled, so it is possible to change the illumination value in manual mode.  

If the Dental Light is turned off, the room light will remain switched. 

 
 

 
 

 
 

 

 
 

 
 

 
 

 

 
 

 
 

 
 

 

 
 

 
 

 
 

 

 
 

 
 

 
 

 
 

 

 
 

 
 

 
 

 

 
 

 
 

 
 

 

Summary of Contents for EVA

Page 1: ......

Page 2: ...LE WHITE VERSION 7 4 4 COMPOSAVE SETTING ON TUNABLE WHITE VERSION 8 4 5 MINIMUM INTENSITY SETTING ON SOLAR VERSION 8 4 6 AUTO ON SETTING 8 4 7 SWITCHING ON OFF THEIA 8 4 8 REMOTE CONTROL 8 4 9 SYNCRO...

Page 3: ...nd the competent authority of the Member State in which the user and or patient is established 1 SYMBOLS USED 1 1 SYMBOLS USED IN THIS MANUAL WARNING The paragraphs marked with this symbol contain ins...

Page 4: ...ce and is intended for illuminating the oral cavity and oral structures of patients in dentistry In the normal use the device is positioned distance of 700mm from the operative area the distance for w...

Page 5: ...DIGIT Fixed Arm 4 DIGIT Light Source Integrated Camera 5 DIGIT Command radiofrequency RF 6 DIGIT Cables 7 DIGIT Custom 1 8 9 DIGIT 5 U 0 Only Head No arms 0 Only Head No arms 0 Tunable White 0 Joystic...

Page 6: ...eters are memorized by the lamp and made available at every switching on Last light intensity setting light colour temperature Setting for Tunable White variant Warning Do not use the device if parts...

Page 7: ...or light changes according to the illuminance level according to the images below Light Intensity Minimum Light Intensity Medium Minimum Light Intensity Medium Maximum Light Intensity Maximum 4 3 CHAN...

Page 8: ...hen the on mode in set in on position the lamps turn on automatically without a specific command from the user when there is power supply 4 7 SWITCHING ON OFF THEIA The light on the fixed arm secondar...

Page 9: ...Light it is necessary to repeat the procedure from point 1 3 After the blue LED switches on the room light there are 60 seconds to confirm the Pairing by pressing the programming button placed on the...

Page 10: ...arter 1 2 must be well secured and safety screws intact The screws under carter 3 must be well secured Yearly Carter 3 Carter 2 Carter 1 Service Engineer Check the absence of any oxidation into joints...

Page 11: ...es Never use detergents containing surfactants or water repellents that depositing can leave streaks Slight streaking will not prejudice the quality of the light Products differing from those suggeste...

Page 12: ...0 60 Hz 32 V dc 10 Max Power 24 V ac 26 VA 32 V dc 14 VA 24 V ac 40 VA 32 V dc 28 VA Recommended fuses for installation not supplied in charge to installation 24 V ac T1 6AL 250V 32 V dc T1 25AL 250V...

Page 13: ...IONS The device in the original packaging can be transported and stored for a maximum period of 15 weeks if the following environmental conditions are met Environmental temperature from 20 C to 70 C R...

Page 14: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...

Reviews: