DESCRIPTION OF COMMON USER’S INTERFACE
1
Joystick
2
Button for pairing and Tk change
3
Sterilizable Handle
4
Knob of the handle
5
Indicator Strip
6
Sensor (alternative to Joystick)
7
Button for Theia Tech
4
INSTRUCTION FOR USE
The device must be cleaned before use (see Device Cleaning paragraph).
Caution
Do not use the device in flammable or explosive environments
Simultaneous use of the light with electro-surgical devices can cause malfunctioning (flickering, no command, etc)
Warning
The Joystick must be handled with care to avoid breakages.
Never move the light using the switch to grip.
Note
At the switching on the device makes a self diagnosis, and the Indicator strip starts blinking with different colours: blue, green
and red.
The following parameters are memorized by the lamp and made available at every switching on:
Last light intensity setting
light colour temperature Setting (for Tunable White variant)
Warning
Do not use the device if parts or enclosures are damaged or plays or gaps appear between:
Head Bow / Head Joint
Fixed Arm Joint / Swivel Arm Joint
Warning
The joystick must be handled with care to avoid breakages.
Never move the light using the joystick.
4.1
SWITCHING ON/ OFF
Joystick
Sensor
Push right or left and release
Acoustic signal: 1 beep
Bring the hand towards the sensor up to 2 cm and move the hand down
Acoustic signal: 1 beep
4.2
ADJUSTING THE LIGHT INSTENSITY
Joystick Increase Light Intensity
Joystick Decrease light intensity
Push left and keep pushed until desired intensity is reached. Then
release
Acoustic signal: 1 beep at command
Maximum intensity reached: continuous beep
Push right and keep pushed until desired intensity is reached, then
release.
Acoustic signal: 1 beep at command
Minimum intensity reached: continuous beep