
EVA
USER MANUAL - EN
Edition 1.0 April 2020
Pag. 10 di 14
5
PREVENTIVE MAINTENANCE AND ROUTINE CHECKS
Only Service Engineer are allowed to perform corrective Maintenance and replacement of any part of the device, according to Manufacturer’s Service
Manual.
Checks
Frequency
Procedure
Responsible
No plays or space gaps between the
junction points (points 1, 2, 3, 4)
Yearly
Service
Engineer
Screws of connection points must
be tightened and integer:
screws 5
Screw 6.
Yearly
Service
Engineer
The nuts of connection points
under carter 1, 2 must be well
secured and safety screws intact.
The screws under carter 3 must be
well secured.
Yearly
Carter 3 Carter 2 Carter 1
Service
Engineer
Check the absence
of any oxidation into joints, arms or
plastic parts.
Yearly
Visual inspection
Service
Engineer
Check the main plate can be read
Yearly
//
Service
Engineer
Check of damages on enclosure and
plastic joints integrity.
Yearly
Service
Engineer
Electrical Safety according EN
62353
1. Dielectric strength
2. Current Leakage
Every two
years
Use the parameters defined into IEC 60601-1
Service
Engineer
Light checks
Every two
years
With a spectroradiometer check the values for:
Max Luminance: >35000 lux
CRI > 85
Radial power on blue light: <100 W/m2
Service
Engineer
Summary of Contents for EVA
Page 1: ......
Page 14: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...