                                      Pag. 10 di 14 

              

5

 

PREVENTIVE MAINTENANCE AND ROUTINE CHECKS 

Only Service Engineer are allowed to perform corrective Maintenance and replacement of any part of the device, according to Manufacturer’s Service 
Manual.   

 

Checks 

Frequency 

Procedure 

Responsible 

No plays or space gaps between the 
junction points (points 1, 2, 3, 4) 

Yearly 

 

 

Service 
Engineer 

Screws of connection points must 

be tightened and integer: 

 

screws 5 

 

Screw 6. 

 
 

Yearly  

    

                               

 

Service 

Engineer 

The nuts of connection points 

under carter 1, 2 must be well 

secured and safety screws intact.  
The screws under carter 3 must be 

well secured. 
 

Yearly 

 

                    Carter 3                        Carter 2                                  Carter 1  

Service 

Engineer 

Check the absence 
of any oxidation into joints, arms or 

plastic parts. 

Yearly 

Visual inspection 

Service 
Engineer 

Check the main plate can be read 

Yearly 

// 

Service 
Engineer 

Check of damages on enclosure and 

plastic joints integrity. 

Yearly 

 

Service 

Engineer 

Electrical Safety according EN 

62353 
1. Dielectric strength 

2. Current Leakage 

Every two 

years 

Use the parameters defined into  IEC 60601-1 

Service 

Engineer 

Light checks 
 

Every two 
years 

With a spectroradiometer check the values for: 
Max Luminance: >35000 lux  

CRI > 85  

Radial power on blue light: <100 W/m2 

Service 
Engineer 

 

 
 

 
 

Summary of Contents for EVA

Page 1: ......

Page 2: ...LE WHITE VERSION 7 4 4 COMPOSAVE SETTING ON TUNABLE WHITE VERSION 8 4 5 MINIMUM INTENSITY SETTING ON SOLAR VERSION 8 4 6 AUTO ON SETTING 8 4 7 SWITCHING ON OFF THEIA 8 4 8 REMOTE CONTROL 8 4 9 SYNCRO...

Page 3: ...nd the competent authority of the Member State in which the user and or patient is established 1 SYMBOLS USED 1 1 SYMBOLS USED IN THIS MANUAL WARNING The paragraphs marked with this symbol contain ins...

Page 4: ...ce and is intended for illuminating the oral cavity and oral structures of patients in dentistry In the normal use the device is positioned distance of 700mm from the operative area the distance for w...

Page 5: ...DIGIT Fixed Arm 4 DIGIT Light Source Integrated Camera 5 DIGIT Command radiofrequency RF 6 DIGIT Cables 7 DIGIT Custom 1 8 9 DIGIT 5 U 0 Only Head No arms 0 Only Head No arms 0 Tunable White 0 Joystic...

Page 6: ...eters are memorized by the lamp and made available at every switching on Last light intensity setting light colour temperature Setting for Tunable White variant Warning Do not use the device if parts...

Page 7: ...or light changes according to the illuminance level according to the images below Light Intensity Minimum Light Intensity Medium Minimum Light Intensity Medium Maximum Light Intensity Maximum 4 3 CHAN...

Page 8: ...hen the on mode in set in on position the lamps turn on automatically without a specific command from the user when there is power supply 4 7 SWITCHING ON OFF THEIA The light on the fixed arm secondar...

Page 9: ...Light it is necessary to repeat the procedure from point 1 3 After the blue LED switches on the room light there are 60 seconds to confirm the Pairing by pressing the programming button placed on the...

Page 10: ...arter 1 2 must be well secured and safety screws intact The screws under carter 3 must be well secured Yearly Carter 3 Carter 2 Carter 1 Service Engineer Check the absence of any oxidation into joints...

Page 11: ...es Never use detergents containing surfactants or water repellents that depositing can leave streaks Slight streaking will not prejudice the quality of the light Products differing from those suggeste...

Page 12: ...0 60 Hz 32 V dc 10 Max Power 24 V ac 26 VA 32 V dc 14 VA 24 V ac 40 VA 32 V dc 28 VA Recommended fuses for installation not supplied in charge to installation 24 V ac T1 6AL 250V 32 V dc T1 25AL 250V...

Page 13: ...IONS The device in the original packaging can be transported and stored for a maximum period of 15 weeks if the following environmental conditions are met Environmental temperature from 20 C to 70 C R...

Page 14: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...

Reviews: