2

The disposal of scrap instrument and its

accessories and packings(including battery, plastic
bags, foams and paper boxes) should follow the
local laws and regulations.

Please check the packing before use to make

sure the device and accessories are totally in
accordance with the packing list, or else the device
may have the possibility of working abnormally.

Please choose the accessories and probe which

are approved or manufactured by the manufacturer,
or else it may damage the device.

Please don't measure this device with functional

tester for the device's related information.

1.3 Attention

Keep the oximeter away from dust, vibration,

corrosive substances, explosive materials, high
temperature and moisture.

If the oximeter gets wet, please stop operating it.

When it is carried from cold environment to

warm or humid environment, please do not use it
immediately.

DO NOT operate keys on front panel with sharp

materials.

High temperature or high pressure steam

disinfection of the oximeter is not permitted. Refer
to User Manual in the relative chapter (7.1)for

3

instructions of cleaning and disinfection.

Do not have the oximeter immerged in liquid.

When it needs cleaning, please wipe its surface with
medical alcohol by soft material. Do not spray any
liquid on the device directly.

When cleaning the device with water, the

temperature should be lower than

60℃.

As to the fingers which are too thin or too cold,

it would probably affect the normal measure of the
patients' SpO

2

and pulse rate, please clip the thick

finger such as thumb and middle finger deeply
enought into the probe.

The update period of data is less than 5 seconds,

which is changeable according to different
individual pulse rate.

Please read the measured value when the

waveform on screen is equably and steady-going,
This measured value is optimal value. And the
waveform at the monment is the standard one.

If some abnormal conditions appear on the

screen during test process, pull out the finger and
reinsert to restore normal use.

The device has normal useful life for three years

since the first electrified use.

The device may not work for all patients. If you

are unable to achieve stable readings, discontinue

Summary of Contents for CMS-50D Plus

Page 1: ......

Page 2: ...the Pulse Oximeter s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as th...

Page 3: ...evice ceaselessly especially for the microcirculation barrier patients It is recommended that the sensor should not be applied to the same finger for over 2 hours For the special patients there should...

Page 4: ...s Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient s safety and monitoring performance about cables and transducers It is recom...

Page 5: ...r is not permitted Refer to User Manual in the relative chapter 7 1 for 3 instructions of cleaning and disinfection Do not have the oximeter immerged in liquid When it needs cleaning please wipe its s...

Page 6: ...able It is only necessary for patients to put one of his fingers into a probe for diagnosis and a display screen will directly show the measured value of pulse oxygen saturation with the high veracity...

Page 7: ...sitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor Figure 1 7 4 Technical specifications 4 1 Main performance A SpO2...

Page 8: ...nm 6 75mW 5 Installation 5 1 View of the front panel Figure 2 front view 5 2 Battery installation A Refer to Figure 3 and insert the two AAA size batteries properly in the right direction 9 B Replace...

Page 9: ...tion The icon of SpO2 is a program for receiving real time data which is shown as figure 5 the icon of SpO2 Review is a program for receiving stored data which is shown as figure 6 11 a Please connect...

Page 10: ...nds and then the screen will be automatically shut down only a flashing yellow dot appear on screen If short press the button at this moment the sign Recording would appear on the screen and then the...

Page 11: ...sensor should not be used at a location 15 or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection E Do not fix the SpO2 sensor with adhesive or else it may result i...

Page 12: ...n shows C Take out the battery if leave the equipment unused for long time D The device needs to be calibrated once a year or according to the calibrating program of hospital It also can be performed...

Page 13: ...vice 1 Please change batteries 2 Please Install the battery again 3 Please contact the local service center The display is off suddenl y 1 The device s malfunction 2 The battery is drained away or alm...

Page 14: ...ion Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED display SpO...

Page 15: ...Moving calculate the Average pulse rate every 4 cardio beat s cycle The deviation between average value and true value does not exceed 1 Safety Type Interior Battery Type 23 Pulse Intensity Range Con...

Page 16: ...24 Dimensions 58 5 L 31 W 32 H mm Weight About 50g with the batteries...

Page 17: ...File Ver 0 9 Release Date March 2009...

Reviews: