16

thiosalicylic hemoglobin, and some with icterus
problem, the SpO

2

determination by this monitor

may be inaccurate.

C

.

The drugs like dopamine, procaine, prilocaine,

lidocaine and butacaine may also be a major factor
blamed for serious error of SpO

2

measure.

D

.

As the SpO

2

value serves as a reference value for

judgement of anemic anoxia and toxic anoxia, some
patients with serious anemia may also report good
SpO

2

measurement.

7

Maintain

transportation and

storage

7.1 Cleaning and Disinfecting

Using medical alcohol to disinfect the device, nature
dry or clean it with clean soft cloth.

7.2 Maintain

A.

Please clean and disinfect the device before using

according to the User Manual(7.1).

B.

Please change the battery when the screen shows

.

C.

Take out the battery if leave the equipment

unused for long time.

D.

The device needs to be calibrated once a year (or

according to the calibrating program of hospital). It
also can be performed at the state-appointed agent or

17

just contact us for calibration.

7.3 Transportation and storage

A.

The packed device can be transported by ordinary

conveyance or according to transport contract. The
device can not be transported mixed with toxic,
harmful, corrosive material . 

B.

The packed device should be stored in room with

no corrosive gases and good ventilation.
Temperature: -40

°

C~60

°

C; Humidity:

95%

8 Troubleshooting

Trouble

Possible

Reason

Solution

The

SpO

2

and

Pulse

Rate
can not

be

displaye
d

normall

y

1. The finger
is not properly
positioned.
2. The
patient

s SpO

2

is too low to
be detected.

1. Place the
finger properly
and try again.
2. Try again;
Go to a
hospital for a
diagnosis if
you are sure
the device
works all right.

Summary of Contents for CMS-50D Plus

Page 1: ......

Page 2: ...the Pulse Oximeter s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as th...

Page 3: ...evice ceaselessly especially for the microcirculation barrier patients It is recommended that the sensor should not be applied to the same finger for over 2 hours For the special patients there should...

Page 4: ...s Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient s safety and monitoring performance about cables and transducers It is recom...

Page 5: ...r is not permitted Refer to User Manual in the relative chapter 7 1 for 3 instructions of cleaning and disinfection Do not have the oximeter immerged in liquid When it needs cleaning please wipe its s...

Page 6: ...able It is only necessary for patients to put one of his fingers into a probe for diagnosis and a display screen will directly show the measured value of pulse oxygen saturation with the high veracity...

Page 7: ...sitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor Figure 1 7 4 Technical specifications 4 1 Main performance A SpO2...

Page 8: ...nm 6 75mW 5 Installation 5 1 View of the front panel Figure 2 front view 5 2 Battery installation A Refer to Figure 3 and insert the two AAA size batteries properly in the right direction 9 B Replace...

Page 9: ...tion The icon of SpO2 is a program for receiving real time data which is shown as figure 5 the icon of SpO2 Review is a program for receiving stored data which is shown as figure 6 11 a Please connect...

Page 10: ...nds and then the screen will be automatically shut down only a flashing yellow dot appear on screen If short press the button at this moment the sign Recording would appear on the screen and then the...

Page 11: ...sensor should not be used at a location 15 or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection E Do not fix the SpO2 sensor with adhesive or else it may result i...

Page 12: ...n shows C Take out the battery if leave the equipment unused for long time D The device needs to be calibrated once a year or according to the calibrating program of hospital It also can be performed...

Page 13: ...vice 1 Please change batteries 2 Please Install the battery again 3 Please contact the local service center The display is off suddenl y 1 The device s malfunction 2 The battery is drained away or alm...

Page 14: ...ion Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED display SpO...

Page 15: ...Moving calculate the Average pulse rate every 4 cardio beat s cycle The deviation between average value and true value does not exceed 1 Safety Type Interior Battery Type 23 Pulse Intensity Range Con...

Page 16: ...24 Dimensions 58 5 L 31 W 32 H mm Weight About 50g with the batteries...