background image

4

use.

2 Overview

The pulse oxygen saturation is the percentage of
HbO

2

in the total Hb in the blood, so-called the O

2

concentration in the blood. It is an important
bio-parameter for the respiration. A number of
diseases relating to respiratory system may cause the
decrease of SpO

2

in the blood, furthermore, some

other causes such as the malfunction of human
body's self-adjustment, damages during surgery, and
the injuries caused by some medical checkup would
also lead to the difficulty of oxygen supply in human
body, and the corresponding symptoms would
appear as a consequence, such as vertigo, impotence,
vomit etc. Serious symptoms might bring danger to
human's life. Therefore, prompt information of
patients' SpO

2

is of great help for the doctor to

discover the potential danger, and is of great
importance in the clinical medical field.
The Pulse Oximeter features in small volume, low
power consumption, convenient operation and being
portable.It is only necessary for patients to put one of
his fingers into a probe for diagnosis, and a display
screen will directly show the measured value of pulse
oxygen saturation with the high veracity and

5

repetition.

2.1 Features
A.

Operation of the product is simple and convenient.

B.

The product is small in volume, light in weight and

convenient in carrying.

C.

Low power consumption

2.2 Major applications and scope of application

The Pulse Oximeter can be used in measuring the
pulse oxygen saturation and pulse rate through finger.
The product is suitable for being used in family,
hospital, oxygen bar, community healthcare, physical
care in sports (It can be used before or after doing
sports, and it is not recommended to use the device
during the process of having sport) and etc.

The problem of overrating would emerge

when the patient is suffering from toxicosis which
caused by carbon monoxide, the device is not
recommended

to

be

used

under

this

circumstance.
2.3 Environment requirements

Storage Environment

a) Temperature :

-40℃~+60℃

 b

) Relative humidity :

5%~95%

 c

) Atmospheric pressure :

500hPa~1060hPa

Operating Environment
a) Temperature:1

0℃~40℃

Summary of Contents for CMS-50D Plus

Page 1: ......

Page 2: ...the Pulse Oximeter s features and requirements main structure functions specifications correct methods for transportation installation usage operation repair maintenance and storage etc as well as th...

Page 3: ...evice ceaselessly especially for the microcirculation barrier patients It is recommended that the sensor should not be applied to the same finger for over 2 hours For the special patients there should...

Page 4: ...s Check the main unit and all accessories periodically to make sure that there is no visible damage that may affect patient s safety and monitoring performance about cables and transducers It is recom...

Page 5: ...r is not permitted Refer to User Manual in the relative chapter 7 1 for 3 instructions of cleaning and disinfection Do not have the oximeter immerged in liquid When it needs cleaning please wipe its s...

Page 6: ...able It is only necessary for patients to put one of his fingers into a probe for diagnosis and a display screen will directly show the measured value of pulse oxygen saturation with the high veracity...

Page 7: ...sitive element information acquired through which will be shown on screen through treatment in electronic circuits and microprocessor Figure 1 7 4 Technical specifications 4 1 Main performance A SpO2...

Page 8: ...nm 6 75mW 5 Installation 5 1 View of the front panel Figure 2 front view 5 2 Battery installation A Refer to Figure 3 and insert the two AAA size batteries properly in the right direction 9 B Replace...

Page 9: ...tion The icon of SpO2 is a program for receiving real time data which is shown as figure 5 the icon of SpO2 Review is a program for receiving stored data which is shown as figure 6 11 a Please connect...

Page 10: ...nds and then the screen will be automatically shut down only a flashing yellow dot appear on screen If short press the button at this moment the sign Recording would appear on the screen and then the...

Page 11: ...sensor should not be used at a location 15 or limb tied with arterial canal or blood pressure cuff or receiving intravenous injection E Do not fix the SpO2 sensor with adhesive or else it may result i...

Page 12: ...n shows C Take out the battery if leave the equipment unused for long time D The device needs to be calibrated once a year or according to the calibrating program of hospital It also can be performed...

Page 13: ...vice 1 Please change batteries 2 Please Install the battery again 3 Please contact the local service center The display is off suddenl y 1 The device s malfunction 2 The battery is drained away or alm...

Page 14: ...ion Specification Information Display Mode The Pulse Oxygen Saturation SpO2 2 digit digital OLED display Pulse Rate PR 3 digit digital OLED display Pulse Intensity bar graph bar graph OLED display SpO...

Page 15: ...Moving calculate the Average pulse rate every 4 cardio beat s cycle The deviation between average value and true value does not exceed 1 Safety Type Interior Battery Type 23 Pulse Intensity Range Con...

Page 16: ...24 Dimensions 58 5 L 31 W 32 H mm Weight About 50g with the batteries...

Page 17: ...File Ver 0 9 Release Date March 2009...

Reviews: