 Features

BATTERY MESSAGING

Charge Battery – For Battery Maintenance

When the battery level gets below 22.7 volts, the display reads ‘CHARGE BATTERY’.

Swap Battery – For Battery Protection

When the battery level gets below 22.3 volts, the display reads ‘SWAP BATTERY’ and only allows the 
unit to go in the down direction.

REMINDER MESSAGING

Inspection Overdue – For Scheduled Maintenance

The purpose of the inspection overdue reminder is to notify the customer that the stand needs to be 
inspected every 30 days.

FAULT MESSAGING

Duty Cycle Fault – For Actuator Protection

The stand actuators are rated for a specific duty cycle. What this means is that for every 20 minutes 
of elapsed time, the stand is rated to run 5 minutes. If the stand is operated in excess of 5 minutes at a 
time, the actuator may be damaged. When the duty cycle fault occurs, the following message comes up: 
‘Actuator Duty Cycle Fault…Cooling’. When this fault occurs, the customer can lower the patient. After 
lowering, the customer will need to wait 15 minutes to cool the actuator motor.

Actuator Overload – For Arm Protection

1.  When an Up or Down command is executed, the circuit board monitors current consumed by the 

actuator. If the actuator consumes too much current, the control will recognize this condition. 

2.  At this time an ‘Actuator Overload’ message will display. The ‘Actuator Overload’ may result from 

lifting a patient/object that is too heavy, arm pinched against obstruction, or mechanical failure 
(internal actuator binding, etc.). 

3.  To override an ‘Actuator Overload’, press the down key. The arm will only operate in the down 

direction during a current overload. Note: This feature is available on all scale and non-scale units.

Weight Overload – For Arm Protection

1.  When an Up or Down command is executed, the control monitors weight readings from the load cell. 

If the weight is measured too high, the control will recognize this condition.

2.  At this time a ‘Weight Overload’ message will display. The Weight Overload may result from 

accidental lifting of a trapped harness or accidental lifting of an overweight patient.  

3.  To override a ‘Weight Overload’, press the down key. The arm will only operate in the down direction 

during a Weight Overload. Note: This feature is only available on scale units.

Summary of Contents for Smart Stand S400

Page 1: ...Service Manual Model S400 and S500 Smart Stand Form 1 145 Rev 02 05 19 WARNING Cancer and Reproductive Harm www P65Warnings ca gov...

Page 2: ...ax 712 542 1899 www ezlifts com Table of Contents Parts Breakdown 3 Monthly Maintenance Checklist 7 Smart Stand Operating Instructions 9 Scale Calibration 10 Advanced Smart Stand Features 11 Wiring Sc...

Page 3: ...Non Scale Parts Diagram EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com Page 3 S400 S500 Non Scale Parts List 400 EZ Smart Stand without Scale Model S400PN Part List 500 EZ Smart Stand witho...

Page 4: ...ok Spring 17 50932 Pad Head Stand 400 lb 18 12013 Actuator Assembly Stand 19 12706 Hand Control Assembly 20 13506 Linkage Rod Assembly Lift Stand 500 EZ Smart Stand w out Scale Model S500PN Item Part...

Page 5: ...With Scale Parts Diagram EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com Page 5 S400 S500 With Scale Parts List 400 EZ Stand with Scale Premier Model 8992 Part List 500 EZ Stand with Scale Pr...

Page 6: ...Stand 400 lb 20 12013 Actuator Assembly Stand 21 12706 Hand Control Assembly 22 13506 Linkage Rod Assembly Lift Stand 500 EZ Smart Stand with Scale Model S500PS Item Part No Description 1 11842 Switch...

Page 7: ...not missing or torn PIN ON FOOT PLATFORM ARM TO MAST BASE PIVOT REAR WHEEL FRONT WHEEL LINKAGE SPREADER EZ Way Smart Stand Safety Maintenance Checklist EZ Way Inc The EZ Way Smart Stand requires a mi...

Page 8: ...Test the Emergency Stop Switch If the switch does not stop the stand when activated call the EZ Way Service Department immediately Perform a functional test of the emergency lowering feature of the s...

Page 9: ...a symbol indicating that the weight is recalled MENUS Enter Customer Settings Menu Weight Calibration 1 After power up sequence press and hold the Down key followed quickly by the Up key and hold bot...

Page 10: ...to ensure the scale is functional 5 Enter the calibration menu After power up sequence press and hold the DOWN key followed quickly by the UP key and hold both keys for 5 seconds 6 Press the UP key to...

Page 11: ...ustomer will need to wait 15 minutes to cool the actuator motor Actuator Overload For Arm Protection 1 When an Up or Down command is executed the circuit board monitors current consumed by the actuato...

Page 12: ...me since the last inspection exceeds 30 days a reminder message will appear when the unit is powered up The purpose of the inspection reminder is to notify the customer that the lift or stand needs to...

Page 13: ...s read only information provides customer feedback for accidental lifting of a trapped harness accidental lifting of arm against obstruction or accidental lifting of overweight patient Lift Stand Tota...

Page 14: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 14 Smart Stand Wiring Schematic...

Page 15: ...ake sure on off button is not flattened out Check if green LED is flashing on board Flashing LED indicates the PCB is getting power Make sure W6 jumper is in place No up down but display works Check c...

Reviews: