EZ Way, Inc.
EZ Way, Inc. • (800) 627-8940 • Fax (712) 542-1899 • www.ezlifts.com
13
Advanced Smart Stand
®
Features
Actuator Minutes – For Tracking Stand Usage
The Actuator Minutes function tracks the accumulated amount of raising and lowering time of the
actuator. With each minute of runtime the ‘Actuator Minutes’ register increments one. A typical lift
takes one minute of runtime. This read-only information provides accurate customer feedback of stand
usage within the care facility. In the event of an actuator replacement the ‘Actuator Minutes’ register can
be reset by calling an EZ Way Service representative.
Actuator Overload Counts – For Diagnostic Tracking with Actuator
The Actuator Overload Counts function tracks Actuator Overload occurrences. When an ‘ACTUATOR
OVERLOAD’ occurs the ‘Actuator Overload Counts’ register increments one. This read-only
information provides customer feedback for; accidental lifting of a trapped harness, accidental lifting of
arm against obstruction, or accidental lifting of overweight patient.
Lift/Stand Total Minutes – For Tracking Stand Usage
‘Lift/Stand Total Minutes’ function tracks the accumulated amount of raising and lowering time of the
actuator on the unit. The ‘Lift/Stand Total Minutes’ is not reset when an actuator is replaced. This read-
only information provides accurate customer feedback of lift usage for the entire life of the stand.
Volts= - For Battery Diagnostics
The ‘Battery Voltage’ function tracks the real-time voltage of the battery. This information is useful in
diagnosing issues related to battery level.
SW Rev – For Software Diagnostics
The ‘Software Revision’ is used for tracking software changes. For example if the original code begins
with version 1.3 and changes are made to the original code the next revision of code will display version
1.4. New controls with version 1.8 will have software updates included.