background image

EZ Way, Inc.

EZ Way, Inc. • (800) 627-8940 • Fax (712) 542-1899 • www.ezlifts.com

10

Smart Stand

®

 Scale Calibration

SCALE CALIBRATION PROCEDURE

1   Nothing should be hanging from the arm assembly or on the foot platform  During the time the 

stand is turned ON and the battery level is displayed, the microprocessor is resetting Zero.

2.  Press the “ON/OFF” key to power up the unit.
3   Pick up or put on footplate an exact and known weight (preferably 200 lb.). (Note: Once in the 

calibration menu, the stand is no longer able to move.)

4.  Press the “WEIGHT” key to ensure the scale is functional.
5.  Enter the calibration menu. After power up sequence, press and hold the “DOWN” key followed 

quickly by the “UP” key and hold both keys for 5 seconds.

6.  Press the “UP” key to calibration menu. Press Enter (right toggle) twice.
7.  Scroll to the proper calibration weight and enter selection. (Note: If harness is not attached, subtract 

3 lb.)

8   “Begin weight calibration” is displayed. Press Enter (right toggle), select Yes, and press Enter (right 

toggle) to execute calibration. Display shows “CALIBRATING”.

9.  When the display leaves this menu selection, it will have been calibrated to the selected weight. Exit 

the calibration menu.

10. Press “WEIGHT” key to verify calibrated weight to 1% accuracy. Calibration is now complete.

Summary of Contents for Smart Stand S400

Page 1: ...Service Manual Model S400 and S500 Smart Stand Form 1 145 Rev 02 05 19 WARNING Cancer and Reproductive Harm www P65Warnings ca gov...

Page 2: ...ax 712 542 1899 www ezlifts com Table of Contents Parts Breakdown 3 Monthly Maintenance Checklist 7 Smart Stand Operating Instructions 9 Scale Calibration 10 Advanced Smart Stand Features 11 Wiring Sc...

Page 3: ...Non Scale Parts Diagram EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com Page 3 S400 S500 Non Scale Parts List 400 EZ Smart Stand without Scale Model S400PN Part List 500 EZ Smart Stand witho...

Page 4: ...ok Spring 17 50932 Pad Head Stand 400 lb 18 12013 Actuator Assembly Stand 19 12706 Hand Control Assembly 20 13506 Linkage Rod Assembly Lift Stand 500 EZ Smart Stand w out Scale Model S500PN Item Part...

Page 5: ...With Scale Parts Diagram EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com Page 5 S400 S500 With Scale Parts List 400 EZ Stand with Scale Premier Model 8992 Part List 500 EZ Stand with Scale Pr...

Page 6: ...Stand 400 lb 20 12013 Actuator Assembly Stand 21 12706 Hand Control Assembly 22 13506 Linkage Rod Assembly Lift Stand 500 EZ Smart Stand with Scale Model S500PS Item Part No Description 1 11842 Switch...

Page 7: ...not missing or torn PIN ON FOOT PLATFORM ARM TO MAST BASE PIVOT REAR WHEEL FRONT WHEEL LINKAGE SPREADER EZ Way Smart Stand Safety Maintenance Checklist EZ Way Inc The EZ Way Smart Stand requires a mi...

Page 8: ...Test the Emergency Stop Switch If the switch does not stop the stand when activated call the EZ Way Service Department immediately Perform a functional test of the emergency lowering feature of the s...

Page 9: ...a symbol indicating that the weight is recalled MENUS Enter Customer Settings Menu Weight Calibration 1 After power up sequence press and hold the Down key followed quickly by the Up key and hold bot...

Page 10: ...to ensure the scale is functional 5 Enter the calibration menu After power up sequence press and hold the DOWN key followed quickly by the UP key and hold both keys for 5 seconds 6 Press the UP key to...

Page 11: ...ustomer will need to wait 15 minutes to cool the actuator motor Actuator Overload For Arm Protection 1 When an Up or Down command is executed the circuit board monitors current consumed by the actuato...

Page 12: ...me since the last inspection exceeds 30 days a reminder message will appear when the unit is powered up The purpose of the inspection reminder is to notify the customer that the lift or stand needs to...

Page 13: ...s read only information provides customer feedback for accidental lifting of a trapped harness accidental lifting of arm against obstruction or accidental lifting of overweight patient Lift Stand Tota...

Page 14: ...EZ Way Inc EZ Way Inc 800 627 8940 Fax 712 542 1899 www ezlifts com 14 Smart Stand Wiring Schematic...

Page 15: ...ake sure on off button is not flattened out Check if green LED is flashing on board Flashing LED indicates the PCB is getting power Make sure W6 jumper is in place No up down but display works Check c...

Reviews: