The treatment timer on the device is used only to support the user. The proper
treatment sequence is in the responsibility of the user. Control the treatment time
with a clock and make a note of potential treatment interruptions in order to ensure
the correct dosage of UV radiation.
2.1.7
Limitation of use
The device must only be used with the included and suitably labeled components,
except requirements made in chapter 2.1.4 and 5.1.2. All components were tested for
the operation of the device and are admitted. Any usage of other components can
result in an undue hazard to the user and the patient. The UV lamp, UV meter and
measurement adapter are carrying serial numbers of each other. This ensures
identification of the matching equipment set.
The device must be used only within the specified conditions.
The device must be used only in designated areas for medical procedures that
conform to national guidelines and standards.
2.1.8
Operating conditions
Temperature
18
–
28 °C
Humidity
max. 70% rel. (not-condensing)
Max. height above SL
2000 m
2.1.9
Storage conditions
Temperature
18
–
28 °C
Additional
Keep dry
2.1.10
Transport conditions
Temperature
-10
–
50 °C
Additional
Keep dry