- 10 -
5.
Installing the cooler and a flask
For an S type glass set
Nozzle set
Nozzle packing
For a vertical cooler (V and E models), carry out “7.
Installing a support bar clamp” on P.13 before
starting the procedures below.
Caution
Be sure to raise the jack before
setting glass parts.
First raise the jack before setting glass parts. If the
jack is raised during setting work, a personal injury
may result.
Caution
Take care for conditions and
handling of glass parts
Broken or flowed glass parts may result in accidents.
Inspect for damages or flows on glass parts and take
care when handling them.
Cooler
Capillary
feed tube
Coolant nozzle
(
gray
)
Vacuum nozzle (white)
Set screw
Housing
Rotary joint
Eyela clip
Specimen flask
Receiver flask
Ball joint clamp
(1) Allow the cooler to tightly contact against the
vacuum seal in the housing and then tighten with
the cap screw.
When you have tighten to some extent, re-tighten
both the cap screw and the cooler at the same time
and position the receiver flask so that its fixing
opening faces downward.
(2) Fix the receiver flask to the cooler with the ball
joint clamp.
(3) Fix the specimen flask to the rotary joint with the
Eyela clip.
(4) Gently insert the capillary into the cooler.
(5) Install two coolant nozzles (gray) and a vacuum
nozzle (white) to the cooler.
Make sure that the nozzle packing sits in the
groove before installation.
(6) Follow the procedures above in the reversed order
to remove.