Operating Manual
6
1.2.2 Class1 Division 1
1.2.2.1 iCAM501U
(Taken from FM approvals certificate of conformity 3042108)
Must be used with Battery specified in this operating manual
Temperature code is T4, Ta=-20°C to 55°C with Duracell Plus or Procell MN1500 batteries or
T3C, Ta=-20°C to 50°C with Duracell Procell PC1500 or Energizer E91 batteries.
1.2.2.2 iCAM502
The iCAM502 is not currently Class1 Division 1 approved.
1.3 List of Notes
The notes supplied in this chapter provide information on the following.
•
Danger! / Warning!
o
Possible hazard to life or health.
•
Caution
o
Possible damage to property.
•
Important
o
Possible damage to enclosure, device or associated equipment.
•
Information
o
Notes on the optimum use of the device
1.3.1 iCAM501U
Warning!
For ATEX applications, ONLY Duracell Plus MN1500 or Duracell Procell
MN1500 cells may be used with the iCAM501U.
Warning!
Any connection to the iCAM501 USB port must NOT be made in a
hazardous area.
Warning!
For Class1 Division 1 applications NOTE THE REDUCED TEMPERATURE
CLASSIFICATION of T4 at 55
o
C when Duracell MN1500/ Duracell Procell
MN1500 cells are used.
Warning!
For Class 1 Division 1 applications ONLY,Duracell Procell PC1500 or
Energizer E91 cells may additionally be used, NOTE THE REDUCED
TEMPERATURE CLASSIFICATION of T3C at 50
o
C ambient when these
cells are used (see Table 1).
Warning!
The user shall NOT mix MN1500, PC1500 or E91 cells in ANY application
Summary of Contents for iCAM50
Page 4: ...Operating Manual 4...
Page 36: ...Operating Manual 36 6 Certification 6 1 iCAM501U EC Declaration of Conformity...
Page 37: ...Operating Manual 37 6 2 iCAM502 EC Declaration of Conformity...
Page 38: ...Operating Manual 38 6 3 iCAM501U Class 1 Division 1 Certificate of Compliance...
Page 39: ...Operating Manual 39...
Page 40: ...Operating Manual 40...
Page 41: ...Operating Manual 41...