![Extech Instruments prc20 User Manual Download Page 7](http://html1.mh-extra.com/html/extech-instruments/prc20/prc20_user-manual_2453744007.webp)
7
PRC20-EU-EN-V1.2 1/13
STORING VALUES INTO MEMORY
There are 5 memory locations for each thermocouple type and for mV. The default values stored in
the these locations are:
Memory
Location
Type
J, K, C, R, S, N
Type
T
Type
E
mV
M1
0.0
o
C
(32.0
o
F)
0.0
o
C
(32.0
o
F)
0.0
o
C
(32.0
o
F)
0mV
M2
100.0
o
C
(212.0
o
F)
100.0
o
C
(212.0
o
F)
100.0
o
C
(212.0
o
F)
10.00mV
M3
500.0
o
C
(932.0
o
F)
200.0
o
C
(392.0
o
F)
200.0
o
C
(392.0
o
F)
25.00mV
M4
750.0
o
C
(1382.0
o
F)
300.0
o
C
(572.0
o
F),
500.0
o
C
(932.0
o
F)
40.00mV
M5
1000.0
o
C
(1832.0
o
F)
400.0
o
C
(752.0
o
F)
750.0
o
C
(1382.0
o
F)
50.00mV
To change the values in memory:
1.
Select the SOURCE mode
2.
Press the SET button to select the memory location to be changed.
3.
Press the arrow buttons to adjust to the new value
4.
Momentarily press the MEM button to store the value. The memory location icon will blink
while the value is being stored..
www.
GlobalTestSupply
.com
Find Quality Products Online at: