1.8 Calibration
Calibration Requirement for EX9017 series. While calibrate type 0D, the EX9017
series need connect external shunt resistor, 125Ohms, 0.1%
Type code
08
09
0A
0B
0C
0D
Zero Input
0V
0V
0V
0mV
0mV
0mA
Span +10V
+5V
+1V
+500mV
+150mV
+20mA
Calibration Sequence:
1. Connect calibration voltage/current to module's channel 0.
2. Warm-Up for 30 minutes
3. Set the input type of module which you wish to calibration.
4. Enable Calibration (P.24)
5. Apply Zero Calibration Voltage
6. Preform Zero Calibration Command (P.16)
7. Apply Span Calibration Voltage
8. Perform Span Calibration Command (P.15)
9. Repeat step4 to step 8 three times.
Warning: Please don't calibrate before you really understand.