15
2.4 $AA0
Description:
Perform Span Calibration
Syntax:
$AA0[CHK](cr)
$ delimiter character
AA address of setting/response module (00 to FF)
0 command for performing zero calibration
Response:
Valid Command:
!AA
Invalid Command:
?AA
Example :
Command : $010 Receive : !01
Perform address 01 zero calibration on channel 0, return success.
Command : $020 Receive : ?02
Perform address 02 zero calibration on channel 2 , return not
enable calibration before perform calibration command.
Warning: Pls don't calibrate before you really understand.
Summary of Contents for EX-9017F